Operator’s Manual ZM-520PA/521PA/530PA/531PA 27
Setting item Description Settings
SpO
2
RESPONSE
Select the SpO
2
response mode.
FAST: Select this for special applications
that require a fast response. FAST is
suitable for detecting short apnea.
NORMAL: For normal monitoring.
SLOW: Select this when you need to suppress a
rapid change in SpO
2
.
FAST, NORMAL,
SLOW
Displaying the PARAMETER SETUP Screen
1. Turn off the transmitter by removing one
battery.
2. While pressing the Function key, turn on the
transmitter (install the battery). The MENU
screen appears.
3. Press the
key to move the cursor to
“PARAMETER SETUP”.
4. Press the Function key to enter PARAMETER
SETUP. The current settings are highlighted.
5. Change settings:
To move the cursor and select the setting
item, press the
or
key then press the
Function key.
To select and register the setting, press the
or
key then press the Function key.
To cancel changing the setting of the
selected item, press the Screen key.
6. When changing settings on the PARAMETER
SETUP screen is complete, press the Screen
key to return to the MENU screen.
7. Press the
or
key to move the cursor to “EXIT”.
8. Press the Function key. The numeric and waveform screen appears.
Cursor
MENU screen
Cursor
MENU screen
Setting item
PARAMETER SETUP screen - page 1
Setting
Cursor
Setting item
PARAMETER SETUP screen - page 1
Setting
Cursor