User's Manual

Operator’s Manual ZM-520PA/521PA/530PA/531PA 31
Changing SYSTEM SETUP Settings
System Setup Setting List
The factory default settings are underlined.
Setting Item Description Settings
PROTOCOL
Select the transmitting protocol.
51: New protocol. A central monitor with ORG-
9100A/9110A multiple patient receiver whose
software version 03-03 or later can receive this
protocol.
41: Old protocol. A central monitor with ORG-
9100A/9110A/9700A multiple patient receiver can
receive this protocol.
NOTE
When 51 is set, the receiving monitor must be able
to receive protocol 51. Otherwise, signals from the
transmitter cannot be received. To use protocol
51, the ORG-9100A/ORG-9110A multiple patient
receiver software version 03-03 or later is required.
51, 41
BRIGHTNESS Select the screen brightness.
DARK,
BRIGHT
FUNCTION KEY
Select the function of the Function key.
SUSPEND ALARM & PAUSE:
Suspends alarm on the receiving monitor for 2
minutes. Pauses monitoring on the transmitter
and receiving monitor.
SUSPEND ALARM:
Suspends alarm on the receiving monitor for 2
minutes.
CONFIRM:
Displays the “PATIENT CONFIRMED” message
on the transmitter screen and transmits the
message to the receiving monitor.
OFF: No function.
NOTE
“SUSPEND ALARM & PAUSE” and “CONFIRM” can
only be set when PROTOCOL is set to 51. To use
protocol 51, the ORG-9100A/ORG-9110A multiple
patient receiver software version 03-03 or later is
required.
SUSPEND
ALARM
& PAUSE,
SUSPEND
ALARM,
CONFIRM,
OFF