User's Manual

32 Operator’s Manual ZM-520PA/521PA/530PA/531PA
Setting Item Description Settings
AUTO RESUME
AFTER PAUSE
Select the interval to resume monitoring after PAUSE.
10 s, 30 s, 1
min, 2 min, 3
min, OFF
Displaying the SYSTEM SETUP Screen
1. Turn off the transmitter by removing one
battery.
2. While pressing the Function key, turn on the
transmitter (insert the battery). The MENU
screen appears.
3. Press the
key to move the cursor to
“SYSTEM SETUP”.
4. Press the Function key to enter SYSTEM
SETUP. The current settings are highlighted.
5. Change settings.
To move the cursor and select the setting
item, press the
or
key then press the
Function key.
To select and register the setting, press the
or
key then press the Function key.
To cancel changing the setting of the
selected item, press the Screen key.
The SYSTEM SETUP screen has two pages.
To display the second page, press the
key
when the cursor is at “BRIGHTNESS”.
6. When changing settings on the SYSTEM
SETUP screen is complete, press the Screen
key to return to the MENU screen.
7. Press the
or
key to move the cursor to
“EXIT”.
8. Press the Function key. The numeric and
waveform screen appears.
Cursor
MENU screen
Cursor
MENU screen
Setting item
SYSTEM SETUP screen - page 1
Setting
Setting item
SYSTEM SETUP screen - page 1
Setting