Operator’s Manual ZM-520PA/521PA/530PA/531PA 19
NOTE
Use Nihon Kohden Medipower to ensure specified performance. Outdated, mismatched
or poor-quality batteries can give unacceptable performance (e.g., insufficient
low battery indication). The use of fresh high quality alkaline batteries is strongly
recommended.
When the display is on, it consumes battery power. Instruct the patient not to turn on
the display during monitoring.
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by the operator. When replacing the batteries
of a transmitter that is currently used for a patient, disconnect the electrode leads
from the transmitter before replacing batteries or do not touch the patient during
replacement.
If electrode leads are attached to the patient and the person replacing batteries touches the patient
GXULQJEDWWHU\UHSODFHPHQWH[FHVVSDWLHQWOHDNDJHFXUUHQWPD\ÀRZLQWRWKHSDWLHQW
NOTE
Replace all batteries at the same time.
Do not use different types of batteries together.
Insert the batteries with the correct polarity (+ and –).
1. Open the battery case cover.
2. Insert two new batteries into the battery case
observing the correct polarity.
The transmitter is automatically turned on
when the batteries are installed.