User's Manual

24 Operator’s Manual ZM-520PA/521PA/530PA/531PA
Turning Off the Transmitter
To turn off the power, remove the batteries. When the power is turned off, the saved waveform and
numeric data are deleted.
Changing the Transmitter Channel
7KHFKDQQHORIWKHWUDQVPLWWHUFDQEHFKDQJHGZLWKWKHRSWLRQDO4,3.FKDQQHOZULWHU
WARNING
The following actions must be taken to properly receive the transmitter signal of the
correct patient on the receiving instrument. Otherwise, there may be signal loss or
signals may mix causing a serious accident, such as monitoring a different patient.
Assign a channel administrator in the hospital and only he or she should manage
channel assignment.
The channel administrator must manage the channels in the facility so that there is
no signal interference.
When the transmitter channel is changed, the channel administrator must check
that the channel on the receiving monitor is also changed and the signal is properly
received.
The channel administrator must replace the channel number label on the
transmitter with the new one after changing the channel.
NOTE
The software version of the QI-901PK channel writer must be 02-01 or later to change
the channel on the transmitter.
The channel writer must be used outside the patient environment.
The channel is displayed in the upper left corner of the screen.
Channel