CAUTION
If the attachment site is dirty with blood
or bodily fluids, clean the attachment
site before attaching the probe. If there
is nail polish on the attachment site,
remove the polish. Otherwise, the
amount of transmitted light decreases,
and measured value may be incorrect
or measurement cannot be performed.
CAUTION
When the probe is attached on an
appropriate site with sufficient
circulation and the error message
confirming the probe attachment
repeatedly appears, the probe may be
deteriorated. Replace it with a new
one.
CAUTION
If the skin gets irritated or redness
appears on the skin from the probe,
change the attachment site or stop
using the probe. Take extreme care for
the patients with delicate skin.
CAUTION
Do not use a probe which is
deteriorated by aging. Accurate
measurement cannot be performed.
CAUTION
Refer to the probe instruction manual
for details.
CAUTION
The disposable probe is not sterilized.
Use the disposable probe only for a
single patient. Never reuse the
disposable probe for another patient
because it causes cross infection.