62 Operator’s Manual ZM-520PA/521PA/530PA/531PA
Use with a Pacemaker
:KHQPRQLWRULQJDSDFHPDNHUSDWLHQWWKHWUDQVPLWWHUGHWHFWVSDFHPDNHUSXOVHDQGUHMHFWVWKH
SDFHPDNHUSXOVHIURPWKHKHDUWUDWHFRXQW
WARNING
The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive
implantable pacemaker may be affected by transmitter which is connected to the
same patient. If this occurs, the pacemaker may pace at its maximum rate and the
transmitter may give incorrect data to the monitor. If this occurs, disconnect the
electrode leads from the patient or change the setting on the pacemaker by referring
to the pacemaker’s manual. For more details, contact your pacemaker representative
or Nihon Kohden representative.
WARNING
The transmitter detects the pacemaker pulse and rejects the pacemaker pulse from
the heart rate count. However, all of the pacemaker pulse might not be rejected. If the
pacemaker pulse is not rejected, the pacemaker pulse is detected as QRS and false
heart rate may be indicated. Keep pacemaker patients under close observation.
* )RUWKHSDFHPDNHUSXOVHUHMHFWLRQFDSDELOLW\RIWKH=03$3$3$3$WUDQVPLWWHU
UHIHUWRWKH³6SHFL¿FDWLRQV(&*´VHFWLRQ
WARNING
The pacemaker pulse can be overlooked or detected as QRS. You cannot confirm the
pacemaker operation only from the detected pacemaker pulse.
NOTE
If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off
the pacing spike detection on the monitor.
Turn the pacing spike detection to ON on the receiving monitor when monitoring a
pacemaker patient. Pacing pulse is detected by the transmitter and transmitted to the
monitor. If the pacing spike detection is turned OFF, QRS and pacemaker spike might
not be distinguished and pacemaker failure might not be recognized.