User's Manual

vi Operator’s Manual ZM-520PA/521PA/530PA/531PA
6. Use with other equipment:
When the equipment and/or system is adjacent to or stacked with other
equipment, the equipment and/or system may affect the other equipment. Before
use, check that the equipment and/or system operates normally with the other
equipment.
7. Use of unspecified accessory, transducer and/or cable:
When an unspecified accessory, transducer and/or cable is connected to this
equipment and/or system, it may cause increased electromagnetic emission
or decreased electromagnetic immunity. The specified configuration of this
equipment and/or system complies with the electromagnetic requirements with
the specified configuration. Only use this equipment and/or system with the
specified configuration.
8. Use of unspecified configuration:
When the equipment and/or system is used with the unspecified system
configuration different than the configuration of EMC testing, it may cause
increased electromagnetic emission or decreased electromagnetic immunity.
Only use this equipment and/or system with the specified configuration.
9. Measurement with excessive sensitivity:
The equipment and/or system is designed to measure bioelectrical signals with
a specified sensitivity. If the equipment and/or system is used with excessive
sensitivity, artifact may appear by electromagnetic interference and this may
cause mis-diagnosis. When unexpected artifact appears, inspect the surrounding
electromagnetic conditions and remove this artifact source.
If the above suggested remedial actions do not solve the problem, consult your
Nihon Kohden representative for additional suggestions.
૞ᚑ㒙ጊޓᖘᏆ ᛚ⹺⌀ᨩޓ⌬