Operator’s Manual ZM-520PA/521PA/530PA/531PA 107
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
- guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m 80 MHz to
2.5 GHz
(1 V/m 80 MHz
to 2.5 GHz for
respiration)
Portable and mobile RF communications
equipment should be used no closer
to any part of the ZM-521PA/531PA,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2
P 80 MHz to 800 MHz
d = 2.3
P 800 MHz to 2.5 GHz
(d = 3.5
P 80 MHz to 800 MHz for
respiration
d = 7.0
P 800 MHz to 2.5 GHz for
respiration)
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as deter mined by an electromagnetic
site survey*
1
, should be less than the
compliance level in each frequency
range*
2
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
*
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the ZM-521PA/531PA is used exceeds the applicable RF compliance level above, the ZM-521PA/531PA
should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the ZM-521PA/531PA.
*
2
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration
and 3 V/m for all other functions.
૞ᚑ㒙ጊޓᖘᏆ ᛚ⹺⌀ᨩޓ⌬