16 Operator’s Manual ZM-520PA/521PA/530PA/531PA
WARNING
When not monitoring SpO
2
, disconnect
the SpO
2
cable from the transmitter.
Otherwise, noise from the probe
sensor may interfere and incorrect data
is displayed on the screen.
CAUTION
Refer to the probe instruction manual
for details.
CAUTION
The disposable probe is not sterilized.
Use the disposable probe only for a
single patient. Never reuse the
disposable probe for another patient
because it causes cross infection.
CAUTION
Do not attach the probe to the same
limb that is used for NIBP
measurement or an IBP catheter. The
SpO
2
measurement may be incorrect.
CAUTION
While a patient is on medication which
causes vasodilation, the pulse
waveform may change and in rare
cases the SpO
2
value might not be
displayed.
CAUTION
Normal external light does not affect
monitoring but strong light such as a
surgical light or sunlight may affect
monitoring. If affected, cover the
measuring site with a blanket.
CAUTION
When a message indicates a faulty
probe, stop monitoring and replace the
probe with a new one.
CAUTION
Do not use a damaged or
disassembled probe. It causes
incorrect measurement and may injure
the patient.
CAUTION
Do not use a probe which is
deteriorated by aging. Accurate
measurement cannot be performed.