User's Manual

20 Operator’s Manual ZM-520PA/521PA/530PA/531PA
NOTE
Use Nihon Kohden Medipower to ensure specified performance. Outdated,
mismatched or poor-quality batteries can give unacceptable performance (e.g.,
insufficient low battery indication). The use of fresh high-quality alkaline batteries is
strongly recommended.
When the display is turned on, it reduces the battery power. Instruct the patient not to
turn on the display during monitoring.
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by the operator. When replacing the
batteries of a transmitter that is currently used for a patient, disconnect the electrode
leads from the transmitter before replacing batteries or do not touch the patient
during replacement.
If electrode leads are attached to the patient and the person replacing batteries touches the patient
during battery replacement, excess patient leakage current may flow into the patient.
NOTE
Replace all batteries at the same time.
Do not use different types of batteries together.
Insert the batteries with the correct polarity (+ and –).
1. Open the battery case cover.
2. Insert two new batteries into the battery case
observing the correct polarity.
The transmitter is automatically turned on
when the batteries are installed.