User's Manual

26 Operator’s Manual ZM-520PA/521PA/530PA/531PA
Changing Parameter and System Setup Settings
The initial settings on the PARAMETER SETUP and SYSTEM SETUP screens can only be
changed before monitoring. Changing these settings during monitoring interrupts monitoring.
NOTE
Changing Parameter and System Setup settings must be done by qualified personnel.
Changing PARAMETER SETUP Settings
Parameter Setup Setting List
The factory default settings are underlined.
Setting item Description Settings
ECG ELECTRODES Select the electrode lead type.
IEC, AHA
LEAD TYPE Select the type of ECG leads.
AUTO, 6 LEADS
ECG
MEASUREMENT
Turn ECG monitoring on or off. When electrodes
are attached to the patient and ECG leads are
connected, ECG monitoring starts even when this
setting is set to OFF.
If this setting is set to OFF, the same setting on the
receiving monitor must also be set to OFF.
NOTE
When “PROTOCOL on the transmitter is set
to 51 and the receiving monitor is able to
receive protocol 51, ECG measurement on
the receiving monitor is automatically set to
OFF when this setting is set to OFF on the
transmitter.
To use protocol 51, the ORG-9100A/ORG-
9110A multiple patient receiver software
version 03-03 or later is required.
ON, OFF
RESP
MEASUREMENT
Turn respiration monitoring on or off.
When this setting is set to OFF, the same setting on
the receiving monitor is automatically set to OFF.
ON, OFF