User's Manual

Operator’s Manual ZM-520PA/521PA/530PA/531PA 35
1. On the SYSTEM SETUP screen, press the
key to move the cursor to “FUNCTION
KEY”. “FUNCTION KEY” is on the second
page of the SYSTEM SETUP screen.
2. Press the Function key.
3. Press the
key to select the function.
4. Press the Function key to register the selected
setting. The cursor returns “FUNCTION
KEY”.
AUTO RESUME AFTER PAUSE
Select the interval to resume monitoring after PAUSE. When one of the following conditions is
met, monitoring resumes on the receiving monitor. When OFF is selected, monitoring does not
automatically resume.
Heart rate is properly monitored for the selected interval
• SpO
2
is properly monitored for the selected interval
1. Press the
key to move the cursor to “AUTO RESUME AFTER PAUSE”. “AUTO
RESUME AFTER PAUSE” is on the second page of the SYSTEM SETUP screen.
2. Press the Function key.
3. Press the
key to select interval.
4. Press the Function key to register the selected setting. The cursor returns to “AUTO
RESUME AFTER PAUSE”.
SYSTEM SETUP screen - page 2SYSTEM SETUP screen - page 2