User's Manual

12 Operator’s Manual ZM-920PA/921PA/930PA/931PA
For Patients Using Implantable Pacemaker
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers
and Cardiac Monitoring and Diagnostic Equipment
The bioelectric impedance measurement sensor of a minute
ventilation rate-adaptive implantable pacemaker may be affected by
transmitter which is connected to the same patient. If this occurs, the
pacemaker may pace at its maximum rate and the transmitter may
give incorrect data to the monitor. If this occurs, disconnect the
electrode leads from the patient or change the setting on the
pacemaker by referring to the pacemaker’s manual. For more details,
contact your pacemaker representative or Nihon Kohden
representative.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the
synchronization signal for other equipment such as IABP, MRI,
echocardiography or defibrillator. There may be time delay between
the monitor and the other equipment caused by waveform
transmission delay and spike noise may interfere on the output signal
and be mistaken as a trigger.