SpO
2
Monitoring
The SpO
2
monitoring is only available on the ZM-930PA/931PA transmitter.
This transmitter sends SpO
2
and pulse waveform to the monitor and displays SpO
2
data and pulse level bar graph on the LCD.
Refer to the operators manual of the monitor for details.
WARNING
SpO
2
measurement may be incorrect in the following cases.
When the patient’s carboxyhemoglobin or methemoglobin
increases abnormally.
When dye is injected in the blood.
When using an electrosurgical unit.
During CPR.
When measuring at a site with venous pulse.
When there is body movement.
When the pulse wave is small (insufficient peripheral circulation).
WARNING
When not monitoring SpO
2
, disconnect the SpO
2
cable from the
transmitter. Otherwise, noise from the probe sensor may interfere and
incorrect data is displayed on the screen.