User's Manual

54 Operator’s Manual ZM-940PA/941PA
Disposable Probes
CAUTION
The disposable probe is not sterilized. Use the disposable probe only for a single
patient. Never reuse the disposable probe for another patient because it causes cross
infection.
Probe Patient Attachment Site
TL-251T Adult
30kgormore
Finger or toe
TL-252T Child
3to40kg
Finger or toe
TL-253T Neonate
3kgorless
Instep and sole
TL-051S/TL-052S
40 mm
Cable length TL-051S: 0.8 m
TL-052S: 1.6 m
Adult
50kgormore
Finger
Neonate
3kgorless
Instep and sole
TL-061S/TL-062S
35 mm
Cable length TL-061S: 0.8 m
TL-062S: 1.6 m
Adult or child
15to50kg
Finger
Infant
3to15kg
Toe