72 Operator’s Manual ZM-940PA/941PA
NIBP
Message Cause Countermeasure
NIBP AIR LEAK
The cuff and extension hose are not
properly connected.
Connect them properly.
The cuff hose (or extension hose) is not
properlyconnectedtotheNIBPsocket.
The cuff or extension hose is damaged. Replace with a new one.
NIBP CANNOT DETECT
PULSE
The patient’s pulse wave is small. Measure by palpation or auscultation.
The cuff is not wrapped on the patient
properly.
Wrap the cuff on the patient properly.
NIBPCHANGE
BATTERIES NO NIBP
NIBP cannot be measured due to low
battery.
Replace batteries with new ones.
NIBP CUFF OCCLUSION Transmitter malfunction.
Immediately remove the cuff from the
patient and contact your Nihon Kohden
representative.
NIBPHIGHCUFFPRESS
Enormous pressure was applied by the
pressure of the cuff.
Remove the cause.
NIBP INFLATION PRESS
LOW
Insufcientcuffinationpressure.
Wait for the remeasurement to be
performedwithincreasedcuffination
pressure.
NIBP MEAS TIME-OUT
The measuring time exceeded the
speciedtimeduetoarrhythmia,body
movement, vibration or, cuff or air hose
being squeezed.
Remove the cause if the cause is body
movement, vibration or squeezing of
cuff or hose.
NIBP MODULE FAILURE Module malfunction.
Contact your Nihon Kohden
representative.
NIBP REMEASURING
NIBP is being remeasured due to
arrhythmia, body movement, vibration
or, cuff or air hose being squeezed.
If the message still appears after
remeasurement, remove the cause if the
cause is body movement, vibration or
squeezing of cuff or hose.
NIBP SAFETY CIRCUIT
RUNNING
(When this message is
displayed, measurement
cannot be performed for 40
seconds.)
Measurement stopped by the safety
circuit.
Checkthatthehoseisnotbentor
squeezed.
Wait 40 seconds, then perform
remeasurement. If the message still
appears, contact your Nihon Kohden
representative.
NIBP SYS OUT OF
RANGE
The maximum blood pressure cannot
be measured even when the cuff
inationpressureexceeded280mmHg
when using adult cuff.
Measure by palpation or auscultation.
NIBP WEAK PULSE
The patient’s pulse wave is too small. Measure by palpation or auscultation.
The cuff is wrapped too loosely. Wrap the cuff properly.
The cuff size is not appropriate. Use the appropriate cuff.
NIBP ZEROING NIBP zero balance is being adjusted.
Do not touch the cuff during zeroing.
Wait for the message to disappear.