User's Manual

Operator’s Manual ZM-940PA/941PA 95
NIBP Measurement
Displayed items: Systolic, diastolic, mean
Cuffpressuredisplayrange: 0to300mmHg
Measurement modes: Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240
minute interval
Measurementaccuracy: ±3mmHg(0≤NIBP≤200mmHg)
±4mmHg(200<NIBP<300mmHg)
Meets or exceeds AAMI Sp-10. 1992 standard
(Maximummeanerror:±5mmHg
Maximumstandarddeviation:8mmHg)
Pulse Rate
Measuring range: 30 to 200 beats/min ±8 beats/min (NIBP)
30 to 250 beats/min ±3% ±1 beat/min (SpO
2
)
Transmitter
FCCregulation: FCCpart95SubpartH
Wireless Medical Telemetry Service (WMTS)
Fieldstrengthlimits: <740mV/m(at3m)
Undesiredemission: below960MHz:200μV/m(at3m)
above960MHz:500μV/m(at3m)
Antenna: Internal
Transmission channel: Indicated on the transmitter
Transmissionfrequencyrange: 1395.0250to1399.9750MHz
1427.0250to1431.9750MHz
Channelspacing: 50kHzor37.5kHz(12.5kHzwheninterleaved)
Modulation: Frequencyshiftkeying
Type of emission: F1D
Occupiedbandwidth: <20kHz
Effective radiated power: 5.0 mW (factory default setting)
Can be changed to 1.0 mW if required
Power Requirements
Operating voltage: 3.2 to 4.8 V
Batterytype: ThreeAA(R6)typeNiMHsecondarybatteries
ThreeAA(R6)typealkalinedrycellprimarybatteries
Battery lifetime:
Type
Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO2 ECG only
NiMHsecondary 1.5 days 2 days 2.5 days
Alkalineprimary 1 day 2 days 2.5 days