WARNING
Never use the transmitter in the
presence of any flammable anesthetic
gas or high concentration oxygen
atmosphere. Failure to follow this
warning may cause explosion or fire.
WARNING
Never use the transmitter in a
hyperbaric oxygen chamber. Failure to
follow this warning may cause
explosion or fire.
WARNING
When performing MRI test, remove all
electrodes from the patient which are
connected to this transmitter. Failure
to follow this warning may cause skin
burn on the patient. For details, refer
to the MRI manual.
WARNING
When performing defibrillation,
discharge as far as possible from
electrodes, patches and any gel,
cream or medicine on the chest of the
patient. If there is a possibility that the
defibrillator paddle could touch these
materials, remove them from the
patient. If the defibrillator paddle
directly contacts these materials, the
discharged energy may cause skin
burn to the patient.
WARNING
Before defibrillation, all persons must
keep clear of the bed and must not
touch the patient or any equipment or
cord connected to the patient. Failure
to follow this warning may cause
electrical shock or injury.
WARNING
When the transmitter is used with an
electrosurgical unit (ESU), firmly
attach the entire area of the ESU
return plate. Otherwise, the current
from the ESU flows into the electrodes
of the transmitter, causing electrical
burn where the electrodes are
attached. For details, refer to the ESU
manual.