12 Operator’s Manual ZM-940PA/941PA
CAUTION
Turn off the power of mobile phones,
small wireless devices and other
devices which produce strong
electromagnetic interference around a
patient (except for devices allowed by
the hospital administrator). Radio
waves from devices such as mobile
phones or small wireless devices may
be mistaken as pulse waves and the
displayed data may be incorrect.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization signal
for other equipment such as IABP, MRI, echocardiography or defibrillator. There may
be time delay between the monitor and the other equipment caused by waveform
transmission delay and spike noise may interfere on the output signal and be
mistaken as a trigger.
Battery
WARNING
Keep the batteries away from fire.
They may explode.
Keep the batteries away from
patients.
Never short-circuit the + and –
terminals on the battery. It may
cause overheating and fire.
Do not damage, disassemble, drop
or give impact to the battery.
WARNING
If the battery is damaged and the
substance inside the battery contacts
the eyes or skin, wash immediately
and thoroughly with water and see a
physician. Never rub your eyes,
because you may lose your eyesight.