Operator's Manual ZS-940PA 5
WARNING
Before performing defibrillation, check that the electrode leads and SpO
2
probe attached to the patient are properly connected to the transmitter.
Touching the metal parts of disconnected leads and probes causes serious
electrical shock or injury by discharged energy.
When performing defibrillation, all persons must keep clear of the bed and
must not touch the patient, any equipment connected to the patient or the
metal parts of leads and probes connected to the patient. Failure to follow this
warning may result in serious electrical burn, shock or other injury.
When performing defibrillation, discharge as far as possible from electrodes
and medicine on the chest of the patient. If there is a possibility that the
defibrillator paddle could touch electrodes and medicine, remove electrodes
and medicine from the patient. If the defibrillator contacts these materials, the
discharged energy may cause serious electrical burn on the patient.
When using this transmitter with an ESU, the ESU return plate and the
electrodes for monitoring must be firmly attached to the patient. If the return
plate is not attached correctly, it may burn the patient’s skin where the
electrodes are attached. Refer to the instruction manual for the ESU.
CAUTION
Do not shake or swing the transmitter holding the leads/cables connected to the
transmitter. The transmitter may come off and injure someone or damage
surrounding instruments.