Operator's Manual ZS-940PA 7
Notes on Parameter Settings
When monitoring NIBP and SpO
2
, the following settings must be set as indicated in the table to
properly transmit the monitoring data to the receiving monitor. Otherwise, data cannot be
displayed on the receiving monitor even when NIBP and SpO
2
are measured on the transmitter.
Some monitors require the software to be upgraded. For details, contact your Nihon Kohden
distributor.
TYPE setting INHIBIT SpO
2
DURING NIBP setting SpO
2
probe attachment site
ON
Probe can be attached to the
same limb as the cuff
A*
1
OFF
Probe must be attached to the
limb without cuff*
2
*
1
Always set to “A”.
*
2
When the SpO
2
probe is attached to the same limb as the NIBP cuff and the cuff is inflated,
the SpO
2
value becomes unstable and SpO
2
or PR alarm may occur.