User's Manual

Operator's Manual ZS-940PA 19
INHIBIT SpO
2
DURING NIBP
Set whether or not to monitor SpO
2
during NIBP measurement.
When the SpO
2
probe is attached to the same limb as the NIBP cuff and this setting is set to
OFF, the pulse may become unstable and SpO
2
or PR alarm may occur. It is recommended to
set this setting to ON so that SpO
2
is not measured during NIBP measurement.
When the SpO
2
probe is attached to the other limb from the NIBP cuff, this setting can be set to
OFF.
NOTE
When the “CHANNEL/TYPE” on the SYSTEM SETUP screen is set to “8”, this
“INHIBIT SpO
2
DURING NIBP” setting becomes invalid and SpO
2
is monitored
during NIBP measurement.
When this “INHIBIT SpO
2
DURING NIBP” is set to OFF, refer to the “Monitoring
SpO
2
during NIBP Measurement” section.
1. Press the NIBP INTERVAL key to move the
cursor to “INHIBIT SpO
2
DURING NIBP”.
2. Press the NIBP START/STOP key to select
“ON” or “OFF”.
ON: Stops SpO
2
monitoring during NIBP measurement.
OFF: SpO
2
is monitored during NIBP measurement.
2ND PARAMETER
Set the display order of SpO
2
and PR.
When set to SpO
2
When set to PR
1. Press the NIBP INTERVAL key to move the
cursor to “2ND PARAMETER”.
2. Press the NIBP START/STOP key to select
“SpO
2
” or “PR”.