68 Operator's Manual ZS-940PA
Transmitter and Electrode Leads
CAUTION
• If detergent or liquid spills into the transmitter, stop using it and contact your
Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in
contact with the transmitter may receive an electric shock or patient leakage
current over the allowed amount may flow.
• Before cleaning or disinfecting the transmitter, remove the batteries from the
transmitter.
• The transmitter cannot be sterilized.
Cleaning
Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or
neutral detergent diluted with water. After cleaning, dry them completely.
Disinfection
CAUTION
• Do not immerse the electrode lead connector in liquid.
• Do not disinfect with hypochlorous acid.
• Use the recommended concentration.
Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth
moistened with any of the disinfectants listed below. Use the recommended concentration.
Disinfectant Concentration (%)
Glutaraldehyde solution 2.0
Hydrochloric alkyl diaminoethylglycine 0.5
Benzalkonium chloride 0.2
Benzethonium chloride solution 0.2
Chlorohexidine gluconate solution 0.5
Cleaning, Disinfection and Sterilization