68 Operator's Manual ZS-940PA
Transmitter and Electrode Leads
CAUTION
If detergent or liquid spills into the transmitter, stop using it and contact your
Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in
contact with the transmitter may receive an electric shock or patient leakage
current over the allowed amount may flow.
Before cleaning or disinfecting the transmitter, remove the batteries from the
transmitter.
The transmitter cannot be sterilized.
Cleaning
Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or
neutral detergent diluted with water. After cleaning, dry them completely.
Disinfection
CAUTION
Do not immerse the electrode lead connector in liquid.
Do not disinfect with hypochlorous acid.
Use the recommended concentration.
Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth
moistened with any of the disinfectants listed below. Use the recommended concentration.
Disinfectant Concentration (%)
Glutaraldehyde solution 2.0
Hydrochloric alkyl diaminoethylglycine 0.5
Benzalkonium chloride 0.2
Benzethonium chloride solution 0.2
Chlorohexidine gluconate solution 0.5
Cleaning, Disinfection and Sterilization