Operator's Manual ZS-940PA 71
NIBP Measurement
Displayed items: Systolic, diastolic, mean
Cuff pressure display range: 0 to 300 mmHg
Measurement modes: Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute
interval
Pulse Rate
Measuring range: 30 to 200 beats/minute ±8 beats/min (NIBP)
30 to 200 beats/min ±3% ±1 beat/min (SpO
2
)
Transmitter
FCC regulation: FCC part 95 Subpart H
Wireless Medical Telemetry Service (WMTS)
Field strength limits: <200 mV/m (at 3 m)
Undesired emission: below 960 MHz: 200 µV/m (at 3 m)
above 960 MHz: 500 µV/m (at 3 m)
Antenna: Internal
Transmission channel: indicated on the transmitter
Transmission frequency range: 608.0125 to 613.9875 MHz
Channel spacing: 25 kHz (12.5 kHz when interleave)
Type of emission: F1D
Occupied bandwidth: <8.5 kHz
Effective radiated power: 1.0 mW (conducted)
Power Requirements
Operating voltage: 3.2 to 4.8 V
Battery type: Three AA type NiMH secondary batteries
Three AA type alkaline dry cell primary batteries
Battery lifetime:
Type Lifetime (Measuring parameters)
ECG, SpO
2
, NIBP ECG, SpO
2
ECG only
NiMH secondary 2 days 2 days 2.5 days
Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is
measured in auto mode at 60 minute intervals and SpO
2
is measured on an index
finger of a male patient with weight 60 kg.
Operation time depends on the thickness of SpO
2
probe attachment site.
Dimension and Weight
Dimension: 114 W × 103 H × 58 D (mm)
Weight: about 350 g (including batteries)