Preparing the LMF – continued Update Antenna Mapping Files Earlier release versions may require the antenna.map file to be updated. There are two antenna mapping files. These are antenna.map and antenna.asu . Follow the steps outlined in Table 3-13 to check the antenna mapping file and update as needed.
Using CDMA LMF Basic LMF Operation NOTE The terms “CDMA LMF” and “WinLMF” are interchangeable 3 The CDMA LMF allows the user to work in the two following operating environments which are accessed using the specified desktop icon: Graphical User Interface (GUI) using the WinLMF icon Command Line Interface (CLI) using the WinLMF CLI icon The GUI is the primary optimization and acceptance testing operating environment.
Using CDMA LMF – continued CLI Format Conventions The CLI command can be broken down in the following way: Verb Device including device identifier parameters Switch Option parameters consisting of: – Keywords 3 – Equals sign (=) between the keyword and the parameter value – Parameter values Spaces are required between the verb, device, switch, and option parameters. A hyphen is required between the device and its identifiers. Following is an example of a CLI command.
Using CDMA LMF – continued BTS Login from the GUI Environment Follow the procedures in Table 3-14 to log into a BTS when using the GUI environment Table 3-14: BTS GUI Login Procedure Step 1 Action Start the LMF GUI environment by double clicking on the WinLMF desktop icon (if the LMF’s not running). 3 NOTE If a warning similar to the following is displayed, select No, shut down other LMF sessions which may be running, and start the LMF GUI environment again: The CLI handler is already running.
Using CDMA LMF – continued Table 3-14: BTS GUI Login Procedure Step 10 Action Click on Login. (A BTS tab with the BTS is displayed.) NOTE If you attempt to log in to a BTS that is already logged on, all devices will be gray. There may be instances where the BTS initiates a log out due to a system error (i.e., a device failure). 3 If the MGLI is OOS_ROM (blue), it will have to be downloaded with code before other devices can be seen.
Using CDMA LMF – continued BTS Login from the CLI Environment Follow the procedures in Table 3-15 to log into a BTS when using the GUI environment Table 3-15: BTS CLI Login Procedure Step 1 Action 3 Double click the WinLMF CLI desktop icon (if the LMF CLI environment is not already running). NOTE If a BTS was logged into under a GUI session when the CLI environment was started, the CLI session will be logged into the same BTS, and step 2 is not required.
Using CDMA LMF – continued Table 3-16: BTS GUI Logout Procedure Step 3 Action Click on Yes or press the Enter key to confirm logout. You are returned to the Login tab. NOTE If a logout was previously performed on the BTS from a CLI window running at the same time as the GUI, a Logout Error popup message will appear stating the system should not log out of the BTS. When this occurs, the GUI must be exited and restarted before it can be used for further operations.
Using CDMA LMF – continued Establishing an MMI Communication Session For those procedures that require MMI communications between the LMF and BTS FRUs, follow the procedure in Table 3-18 to initiate the communication session. Table 3-18: Establishing MMI Communications Step Action 3 1 Connect the LMF computer to the equipment as detailed in the applicable procedure that requires MMI communication session.
Download the BTS Overview Before a BTS can operate, each equipped device must contain device initialization (ROM) code. ROM code is loaded in all devices during manufacture, factory repair, or, for software upgrades, from the CBSC using the DownLoad Manager (DLM). Device application (RAM) code and data must be downloaded to each equipped device by the user before the BTS can be made fully functional for the site where it is installed.
Download the BTS – continued When code is downloaded to an MGLI or GLI, the LMF automatically also downloads data and then enables the MGLI. When enabled, the MGLI will change to INS_ACT (bright green). A redundant GLI will not be automatically enabled and will remain OOS_RAM (yellow). When the redundant GLI is manually commanded to enable through the LMF, it will change state to INS_SBY (olive green). For non–GLI devices, data must be downloaded after RAM code is downloaded.
Download the BTS – continued Table 3-19: Verify GLI ROM Code Loads Step 3 Action 3 In the status report window which opens, note the number in the ROM Ver column for each GLI2. 4 If the ROM code loaded in the GLIs is not the correct one for the software release being used on the BSS, log out of the BTS, disconnect the LMF computer, reconnect the span lines as described in Table 5-6, and have the CBSC download the correct ROM code version to the BTS devices.
Download the BTS – continued Table 3-20: Download and Enable MGLI and GLI Devices Step Action 5 Once the MGLI is enabled, load and enable additional installed GLIs by clicking on the devices and repeating steps 3 and 4. 6 Click OK to close the status window for the additional GLI devices. 3 Download RAM Code and Data to Non–GLI Devices Downloads to non–GLI devices can be performed individually for each device or all installed devices can be downloaded with one action.
Download the BTS – continued Table 3-21: Download RAM Code and Data to Non–GLI Devices Step 5 6 3 Action Click Device in the BTS menu bar, and select select Download > Data in the pull–down menus. – A status report is displayed showing the results of the download for each selected device. Click OK to close the status report window when downloading is completed. Select CSM Clock Source A CSM can one of have three different clock sources.
Download the BTS – continued on–board GPS module (RF–GPS) or a remote GPS receiver (R–GPS). The CSM2 card is required when using the R–GPS. The GPS receiver (mounted on CSM 1) is used as the primary timing reference and synchronizes the entire cellular system. CSM 2 provides redundancy (but does not have a GPS receiver). The BTS may be equipped with a LORAN–C LFR, HSO, or external 10 MHz Rubidium source which the CSM can use as a secondary timing reference. The HSOX is used for expansion frames.
Download the BTS – continued Table 3-23: Enable CSMs Step Action NOTE 2 3 If equipped with two CSMs, CSM–1 should be bright green (INS–ACT) and CSM–2 should be dark green (INS–STY) If more than an hour has passed, refer to CSM Verification, see Figure 3-8 and Table 3-26 to determine the cause. NOTE After the CSMs have been successfully enabled, observe the PWR/ALM LEDs are steady green (alternating green/red indicates the card is in an alarm state).
CSM System Time – GPS & LFR/HSO Verification Clock Synchronization Manager System Time The primary function of the Clock Synchronization Manager (CSM) boards (slots 1 and 2) is to maintain CDMA system time. The CSM in slot 1 is the primary timing source while slot 2 provides redundancy. The CSM2 card (CSM second generation) is required when using the remote GPS receiver (R–GPS). R–GPS uses a GPS receiver in the antenna head that has a digital output to the CSM2 card.
CSM System Time – GPS & LFR/HSO Verification – continued The HSO is a high stability 10 MHz oscillator with the necessary interface to the CSMs. The HSO is typically installed in those geographical areas not covered by the LORAN–C system. Since the HSO is a free–standing oscillator, system time can only be maintained for 24 hours after 24 hours of GPS lock. Upgrades and Expansions: LFR2/HSO2/HSOX LFR2/HSO2 (second generation cards) both export a timing signal to the expansion frames.
CSM System Time – GPS & LFR/HSO Verification – continued CSM Frequency Verification The objective of this procedure is the initial verification of the CSM boards before performing the RF path verification tests. Parts of this procedure will be repeated for final verification after the overall optimization has been completed. Test Equipment Setup (GPS & LFR/HSO Verification) 3 Follow the steps outlined in Table 3-25 to set up test equipment.
CSM System Time – GPS & LFR/HSO Verification – continued Figure 3-8: CSM MMI Terminal Connection REFERENCE OSCILLATOR CSM board shown removed from frame MMI SERIAL PORT 3 EVEN SECOND TICK TEST POINT REFERENCE GPS RECEIVER ANTENNA INPUT ANTENNA COAX CABLE GPS RECEIVER 19.6 MHZ TEST POINT REFERENCE (NOTE 1) NULL MODEM BOARD (TRN9666A) 9–PIN TO 9–PIN RS–232 CABLE FW00372 LMF NOTEBOOK DB9–TO–DB25 ADAPTER COM1 NOTES: 1.
CSM System Time – GPS & LFR/HSO Verification – continued Table 3-26: GPS Initialization/Verification Step Action 1 To verify that Clock alarms (0000), Dpll is locked and has a reference source, and GPS self test passed messages are displayed within the report, issue the following MMI command bstatus – Observe the following typical response: 3 CSM Status INS:ACTIVE Slot A Clock MASTER. Clock Alarms (0000): DPLL is locked and has a reference source.
CSM System Time – GPS & LFR/HSO Verification – continued Table 3-26: GPS Initialization/Verification Step Action 3 HSO information (underlined text above, verified from left to right) is usually the #1 reference source. If this is not the case, have the OMCR determine the correct BTS timing source has been identified in the database by entering the display bts csmgen command and correct as required using the edit csm csmgen refsrc command.
CSM System Time – GPS & LFR/HSO Verification – continued Table 3-26: GPS Initialization/Verification Step 5 Action Enter the following command at the CSM> prompt to verify that the GPS receiver is in tracking mode.
CSM System Time – GPS & LFR/HSO Verification – continued Table 3-26: GPS Initialization/Verification Step 7 Action If steps 1 through 6 pass, the GPS is good. * IMPORTANT If any of the above mentioned areas fail, verify that: – If Initial position accuracy is “estimated” (typical), at least 4 satellites must be tracked and visible (1 satellite must be tracked and visible if actual lat, log, and height data for this site has been entered into CDF file).
CSM System Time – GPS & LFR/HSO Verification – continued LORAN–C Initialization/Verification Table 3-27: LORAN–C Initialization/Verification Step Action 1 At the CSM> prompt, enter lstatus to verify that the LFR is in tracking mode.
CSM System Time – GPS & LFR/HSO Verification – continued Table 3-27: LORAN–C Initialization/Verification Step Action Note 2 Verify the following LFR information (highlighted above in boldface type): – Locate the “dot” that indicates the current phase locked station assignment (assigned by MM). – Verify that the station call letters are as specified in site documentation as well as M X Y Z assignment. – Verify the S/N ratio of the phase locked station is greater than 8.
Test Equipment Set–up Connecting Test Equipment to the BTS The following test equipment is required to perform calibration and ATP tests: LMF Communications system analyzer model supported by the LMF Power meter model supported by the LMF (required when using the HP 8921A/600 and Advantest R3465 analyzers) Non–radiating transmit line termination load Directional coupler and in–line attenuator RF cables and adapters Refer to Table 3-28 and Table 3-29 for an overview of connections for test equip
Test Equipment Set–up – continued IS–95 A/B Testing Optimization and ATP testing for IS–95A/B sites or carriers may be performed using one of the following test equipment: CyberTest Advantest R3267 spectrum analyzer with R3562 signal generator and HP–437B or Gigatronics Power Meter Agilent E4406A transmitter test set with E4432B signal generator Agilent 8935 series E6380A communications test set (formerly HP 3 8935) Hewlett–Packard HP 8921 (with CDMA interface for 1.
Test Equipment Set–up – continued Test Equipment Connection Charts To use the following charts to identify necessary test equipment connections, locate the communications system analyzer being used in the COMMUNICATIONS SYSTEM ANALYZER columns, and read down the column. Where a dot appears in the column, connect one end of the test cable to that connector. Follow the horizontal line to locate the end connection(s), reading up the column to identify the appropriate equipment and/or BTS connector.
Test Equipment Set–up – continued CDMA2000 1X/IS–95A/B–capable Test Equipment Connections Table 3-29 depicts the interconnection requirements for currently available test equipment supporting both CDMA 2000 1X and IS–95A/B which meets Motorola standards and is supported by the LMF. Table 3-29: CDMA2000 1X/IS–95A/B Test Equipment Interconnection COMMUNICATIONS SYSTEM ANALYZER 3 SIGNAL EVEN SECOND SYNCHRONIZATION 19.
Test Equipment Set–up – continued Equipment Warm-up IMPORTANT * Warm-up BTS equipment for a minimum of 60 minutes prior to performing the BTS optimization procedure. This assures BTS stability and contributes to optimization accuracy. – Time spent running initial or normal power-up, hardware/firmware audit, and BTS download counts as warm-up time. WARNING Before installing any test equipment directly to any BTS TX OUT connector, verify there are no CDMA channels keyed.
Test Equipment Set–up – continued Figure 3-9: Cable Calibration Test Setup – CyberTest, Agilent 8935, Advantest R3465, and HP 8921A SUPPORTED TEST SETS CALIBRATION SET UP Motorola CyberTest A. SHORT CABLE CAL ÏÏÏ ÏÏÏÌ 3 ANT IN SHORT CABLE TEST SET RF GEN OUT Note: The 30 dB directional coupler is not used with the Cybertest test set. The TX cable is connected directly to the Cybertest test set. B.
Test Equipment Set–up – continued Figure 3-10: Cable Calibration Test Setup – Agilent E4406A/E4432B and Advantest R3267/R3562 SUPPORTED TEST SETS CALIBRATION SET UP A. SHORT CABLE CAL Agilent E4432B (Top) and E4406A (Bottom) SHORT CABLE RF OUTPUT 50 Ω TEST SET 3 B. RX TEST SETUP FOR TRDC N–N FEMALE ADAPTER RF INPUT 50 Ω RX CABLE SHORT CABLE NOTE: TEST SET 10 MHZ IN ON REAR OF SIGNAL GENERATOR IS CONNECTED TO 10 MHZ OUT (SWITCHED) ON REAR OF TRANSMITTER TESTER (FIGURE F-5).
Test Equipment Set–up – continued Set–up for TX Calibration Figure 3-11 and Figure 3-12 show the test set connections for TX calibration. Figure 3-11: TX Calibration Test Setup – CyberTest (IS–95A/B) and Agilent 8935 (IS–95A/B and CDMA2000 1X) TEST SETS TRANSMIT (TX) SET UP 3 Motorola CyberTest ÏÏÏ ÏÏÏ ÏÏÏÌ FRONT PANEL POWER SENSOR NOTE: IF BTS IS EQUIPPED WITH DRDCS (DUPLEXED RX/TX SIGNALS), CONNECT THE TX TEST CABLE TO THE DRDC ANTENNA CONNECTOR. COMMUNICATIONS TEST SET 100–WATT (MIN.
Test Equipment Set–up – continued Figure 3-12: TX Calibration Test Setup – Using Power Meter TEST SETS TRANSMIT (TX) SET UP NOTE: THE HP8921A AND ADVANTEST R3465 CANNOT BE USED FOR TX CALIBRATION. A POWER METER MUST BE USED. POWER SENSOR NOTE: IF BTS IS EQUIPPED WITH DRDCS (DUPLEXED RX/TX SIGNALS), CONNECT THE TX TEST CABLE TO THE DRDC ANTENNA CONNECTOR. POWER METER 3 100–WATT (MIN.) NON–RADIATING RF LOAD TX TEST CABLE DIRECTIONAL COUPLER (30 DB) 50 Ω TERM .
Test Equipment Set–up – continued Figure 3-13: TX Calibration Test Setup – Agilent E4406A and Advantest R3567 (IS–95A/B and CDMA2000 1X) TEST SETS TRANSMIT (TX) SET UP Agilent E4406A POWER SENSOR NOTE: IF BTS IS EQUIPPED WITH DRDCS (DUPLEXED RX/TX SIGNALS), CONNECT THE TX TEST CABLE TO THE DRDC ANTENNA CONNECTOR. 3 COMMUNICATIONS TEST SET RF INPUT 50 Ω OR INPUT 50 Ω 100–WATT (MIN.
Test Equipment Set–up – continued Set–up for Optimization/ATP Figure 3-14 and Figure 3-15 show the test set connections for optimization/ATP tests. Figure 3-14: IS–95A/B Optimization/ATP Test Set–up, TRDC Shown – CyberTest and Advantest R3465 TEST SETS Optimization/ATP SET UP Motorola CyberTest SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD RF GEN OUT OR RF OUT 50Ω RX TEST CABLE FREQ MONITOR 19.
Test Equipment Set–up – continued Figure 3-15: IS–95A/B Optimization/ATP Test Setup – HP 8921A TEST SETS Optimization/ATP SET UP RX TEST CABLE Hewlett Packard Model HP 8921A W/PCS Interface (for 1900 MHz) SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD RF OUT ONLY NOTE: IF BTS IS EQUIPPED WITH DRDCS (DUPLEXED RX/TX SIGNALS), BOTH THE TX AND RX TEST CABLES CONNECT TO THE DRDC ANTENNA CONNECTOR. (SEE FIGURE 3-16.) FREQ MONITOR 19.
Test Equipment Set–up – continued Figure 3-16: IS–95A/B and CDMA2000 1X Optimization/ATP Test Setup With DRDCs – Agilent Test Equipment TEST SETS Optimization/ATP SET UP RF OUTPUT 50 Ω OR DUPLEX OUT Agilent 8935 Series E6380A (formerly HP 8935) SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD 10 MHZ IN HP–IB TO GPIB BOX RF IN/OUT OR RF INPUT 50 Ω ÁÁ ÁÁ ÁÁ ÁÁ DUPLEX OUT PATTERN TRIG IN RX TEST CABLE FREQ MONITOR 19.6608 MHZ CLOCK REFERENCE FROM CSM BOARD SIGNAL GENERATOR 100–WATT (MIN.
Test Equipment Set–up – continued Figure 3-17: IS–95A/B and CDMA2000 1X Optimization/ATP Test Setup With DRDCs – Advantest R3267/3562 Test Equipment TEST SETS Optimization/ATP SET UP RF OUT 50 Ω Advantest R3267 (Top) and R3562 (Bottom) SIGNAL GENERATOR MOD TIME BASE IN SYNTHE REF IN EXT TRIG IN RX TEST CABLE TO EXT TRIG ON REAR OF SPECTRUM ANALYZER GPIB SPECTRUM ANALYZER 3 100–WATT (MIN.
Test Equipment Set–up – continued Figure 3-18: IS–95A/B and CDMA2000 1X Optimization/ATP Test Setup With TRDCs – Agilent Test Equipment TEST SETS Optimization/ATP SET UP SYNC MONITOR EVEN SEC TICK PULSE REFERENCE FROM CSM BOARD FREQ MONITOR 19.6608 MHZ CLOCK REFERENCE FROM CSM BOARD SIGNAL GENERATOR 10 MHZ IN PATTERN TRIG IN HP–IB TO GPIB BOX RF IN/OUT OR RF INPUT 50 Ω ÁÁ ÁÁ ÁÁ ÁÁ DUPLEX OUT RF OUTPUT 50 Ω OR DUPLEX OUT RX TEST CABLE Agilent 8935 Model E6380A (formerly HP 8935) 100–WATT (MIN.
Test Equipment Set–up – continued Figure 3-19: IS–95A/B and CDMA2000 1X Optimization/ATP Test Setup With TRDCs – Advantest R3267/3562 Test Equipment TEST SETS Optimization/ATP SET UP RF OUT 50 Ω RX TEST CABLE Advantest R3267 (Top) and R3562 (Bottom) SIGNAL GENERATOR MOD TIME BASE IN TO EXT TRIG ON REAR OF SPECTRUM ANALYZER 3 SYNTHE REF IN EXT TRIG IN GPIB SPECTRUM ANALYZER 100–WATT (MIN.
Test Equipment Set–up – continued TX ATP Setup Figure 3-20 shows a typical TX ATP setup.
Test Equipment Set–up – continued Figure 3-21: Typical RX ATP Setup with Directional Coupler (shown with or without RFDS) COBRA RFDS Detail RX ANTENNA DIRECTIONAL COUPLERS RX RF FROM BTS FRAME 2 3 4 5 6 RX (RFM TX) 1 TX (RFM RX) RFDS TX (RFM RX) COUPLER OUTPUTS TO RFDS FWD(BTS) ASU1 (SHADED) CONNECTORS 3 RF FEED LINE TO TX ANTENNA REMOVED Connect RX test cable between the test set and the appropriate RX antenna directional coupler.
Test Set Calibration Background Proper test equipment setup ensures that the test equipment and associated test cables do not introduce measurement errors, and that measurements are correct. NOTE If the test set being used to interface with the BTS has been calibrated and maintained as a set, this procedure does not need to be performed. (Test Set includes LMF terminal, communications test set, additional test equipment, associated test cables, and adapters.
Test Set Calibration – continued GPIB Addresses GPIB addresses can range from 1 through 30. The LMF will accept any address in that range, but the numbers entered in the LMF Options window GPIB address box must match the addresses of the test equipment. Motorola recommends using 1 for a CDMA signal generator, 13 for a power meter, and 18 for a communications system analyzer.
Test Set Calibration – continued Table 3-30: Selecting Test Equipment Manually in a Serial Connection Tab Step 6 Action Type the GPIB address in the corresponding GPIB address box (refer to the Setting GPIB Addresses section of Appendix F for directions on verifying and/or changing test equipment GPIB addresses).
Test Set Calibration – continued Automatically Selecting Test Equipment in a Serial Connection Tab When using the auto-detection feature to select test equipment, the CDMA LMF examines which test equipment items are actually communicating with CDMA LMF. Follow the procedure in Table 3-31 to use the auto-detect feature. Table 3-31: Selecting Test Equipment Using Auto-Detect 3 Step Action 1 From the Tools menu, select Options. The LMF Options window appears.
Test Set Calibration – continued Calibrating Test Equipment The calibrate test equipment function zeros the power measurement level of the test equipment item that is to be used for TX calibration and audit. If both a power meter and an analyzer are connected, only the power meter is zeroed. Calibrate Test Equipment from the Util menu list is used to calibrate test equipment item before being used for testing. The test equipment must be selected before beginning calibration.
Test Set Calibration – continued Calibrating Cables with a CDMA Analyzer The Cable Calibration menu item from the Util menu list is used to calibrate both TX and RX test cables for use with CDMA LMF. NOTE LMF cable calibration cannot be accomplished with an HP8921A analyzer for 1.9 MHz. A different analyzer type or the signal generator and spectrum analyzer method must be used (refer to Table 3-34 and Table 3-35).
Test Set Calibration – continued Calibrating TX Cables Using a Signal Generator and Spectrum Analyzer Follow the procedure in Table 3-34 to calibrate the TX cables using the signal generator and spectrum analyzer. Refer to Figure 3-22 for a diagram of the signal generator and spectrum analyzer. Table 3-34: Calibrating TX Cables Using Signal Generator and Spectrum Analyzer Step 3 Action 1 Connect a short test cable between the spectrum analyzer and the signal generator.
Test Set Calibration – continued Calibrating RX Cables Using a Signal Generator and Spectrum Analyzer Follow the procedure in Table 3-35 to calibrate the RX cables using the signal generator and spectrum analyzer. Refer to Figure 3-23, if required. Table 3-35: Calibrating RX Cables Using a Signal Generator and Spectrum Analyzer Step 3 Action 1 Connect a short test cable to the spectrum analyzer and connect the other end to the Signal Generator.
Test Set Calibration – continued Setting Cable Loss Values Cable loss values for the TX and RX test cable configurations are normally set by accomplishing cable calibration with use of the applicable test equipment. The resulting values are stored in the cable loss files. The cable loss values can also be set/changed manually. Prerequisites Logged into the BTS 3 Table 3-36: Setting Cable Loss Values Step Action 1 Click on the Util menu. 2 Select Edit >Cable Loss > TX or RX.
Test Set Calibration – continued Setting Coupler Loss Value If an in–service coupler is installed the coupler loss (e.g., 30 dB) must be manually entered so it will be included in the LMF TX calibration and audit calculations and the RX FER test. Prerequisites Logged into the BTS Table 3-37: Setting Coupler Loss Values 3 Step Action 1 Click on the Util menu. 2 Select Edit >Coupler Loss>TX or RX. A data entry pop–up window will appear.
Bay Level Offset Calibration Introduction Calibration compensates for normal equipment variations within the BTS and assures maximum measurement accuracy. RF Path Bay Level Offset Calibration Calibration identifies the accumulated gain in every transmit path (BBX slot) at the BTS site and stores that value in the CAL file. The BLOs are subsequently downloaded to each BBX. Each receive path starts at a BTS RX antenna port and terminates at a backplane BBX slot.
Bay Level Offset Calibration – continued TX Path Calibration The TX Path Calibration assures correct site installation, cabling, and the first order functionality of all installed equipment. The proper function of each RF path is verified during calibration. The external test equipment is used to validate/calibrate the TX paths of the BTS. WARNING Before installing any test equipment directly to any TX OUT connector you must first verify that there are no CDMA channels keyed.
Bay Level Offset Calibration – continued BLO Calibration Data File During the calibration process, the LMF creates a calibration (BLO) data file. After calibration has been completed, this offset data must be downloaded to the BBXs using the Download BLO function. An explanation of the file is shown below. NOTE Due to the size of the file, Motorola recommends that you print out a hard copy of a bts.cal file and refer to it for the following descriptions.
Bay Level Offset Calibration – continued – The second breakdown of the array is per sector. Three sectors are allowed. Table 3-39: BTS.
Bay Level Offset Calibration – continued The 20 calibration entries for each slot/branch combination must be stored in order of increasing frequency. If less than 10 points (frequencies) are calibrated, the largest frequency that is calibrated is repeated to fill out the 10 points. Example: C[1]=384, odd cal entry = 1 ‘‘calibration point” C[2]=19102, even cal entry C[3]=777, C[4]=19086, . .
Bay Level Offset Calibration – continued Transmit (TX) Path Calibration The assigned channel frequency and power level (as measured at the top of the frame) for transmit calibration is derived from the site CDF file. For each BBX, the channel frequency is specified in the ChannelList CDF file parameter and the power is specified in the SIFPilotPwr CDF file parameter for the sector associated with the BBX (located under the ParentSECTOR field of the ParentCARRIER CDF file parameter).
Bay Level Offset Calibration – continued CDF – This pattern setting is for advanced users who need to use CDF gain settings for all channels included in the Standard pattern setting (pilot, paging, synch, and six traffic). Using this pattern setting requires the selection of both a BBX and at least one MCC. Verify BLO In both the TX Calibration and All Cal/Audit dialog boxes, a Verify BLO checkbox is provided and checked by default.
Bay Level Offset Calibration – continued IMPORTANT * 3 Verify all BBX boards removed and repositioned have been returned to their assigned shelves/slots. Any BBX boards moved since they were downloaded will have to be downloaded again. Table 3-41: BTS TX Path Calibration Step 1 Action Select the BBX(s) to be calibrated. NOTE If STANDARD, CDF or CDFPILOT is selected for TEST PATTERN, then at least one MCC must be also selected. 2 From the Tests menu, select TX>TX Calibration.
Bay Level Offset Calibration – continued Exception Handling In the event of a failure, the calibration procedure displays a FAIL message in the status report window and provides information in the Description field. Recheck the test setup and connection and re–run the test. If the tests fail again, note specifics about the failure, and refer to Chapter 6, Troubleshooting. 3 Download BLO Procedure After a successful TX path calibration, download the BLO calibration file data to the BBXs.
Bay Level Offset Calibration – continued Calibration Audit Introduction The BLO calibration audit procedure confirms the successful generation and storage of the BLO calibrations. The calibration audit procedure measures the path gain or loss of every BBX transmit path at the site. In this test, actual system tolerances are used to determine the success or failure of a test. The same external test equipment set up is used.
Bay Level Offset Calibration – continued TX Audit Test The Tests menu item, TX Audit, performs the TX BLO Audit test for a BBX(s). All measurements are made through the appropriate TX output connector using the calibrated TX cable setup. Prerequisites Before running this test, the following should be done: CSM–1,GLI2s, BBXs have correct code load. 3 Primary CSM and MGLI2 are INS. All BBXs are OOS_RAM. Test equipment and test cables are calibrated and connected for TX BLO calibration.
Bay Level Offset Calibration – continued Table 3-43: BTS TX Path Audit Step 1 Action Select the BBX(s) to be audited. NOTE If STANDARD or CDF is selected for Test Pattern, then at least one MCC must be also selected. 3 2 From the Tests menu, select TX>TX Audit. 3 4 Select the appropriate carrier(s) displayed in the Channels/Carrier pick list. Press and hold the or key to select multiple items. Type the appropriate channel number in the Carrier n Channels box.
Bay Level Offset Calibration – continued All Cal/Audit Test The Tests menu item, All Cal/Audit, performs the TX BLO Calibration and Audit test for a XCVR(s). All measurements are made through the appropriate TX output connector using the calibrated TX cable setup. NOTE If the TX calibration portion of the test passed, the BLO data will automatically be downloaded to the BBX(s) before the audit portion of the test is run.
Bay Level Offset Calibration – continued Table 3-44: All Cal/Audit Test Step 1 Action Select the BBX(s) to be tested. NOTE If STANDARD, CDF or CDFPILOT is selected for TEST PATTERN, then at least one MCC must be also selected. 3 2 From the Tests menu, select All Cal/Audit. 3 4 Select the appropriate carrier(s) displayed in the Channels/Carrier pick list. Press and hold the or key to select multiple items. Type the appropriate channel number in the Carrier n Channels box.
Bay Level Offset Calibration – continued Create CAL File The Create Cal File function gets the BLO data from BBXs and creates/updates the CAL file for the BTS. If a CAL file does not exist a new one is created. If a CAL file already exists it is updated. After a BTS has been fully optimized a copy of the CAL file must exist so it can be transferred to the CBSC. If TX calibration has been successfully performed for all BBXs and BLO data has been downloaded, a CAL file will exist.
RFDS Setup and Calibration RFDS Description The optional RFDS is used to perform RF tests of the site from the CBSC or from the LMF. The RFDS contains the following FRUs: Antenna Select Unit (ASU) Fixed Wireless Terminal Interface Card (FWTIC) Subscriber Unit Assembly (SUA) For complete information regarding the RFDS, refer to the CDMA CDMA RFDS Hardware Installation; 68P64113A93, CDMA RFDS User’s Guide; 68P64114A51, and the LMF Help function on–line documentation. 3 RFDS Parameters The bts–#.
RFDS Setup and Calibration – continued Table 3-46: RFDS Parameter Settings Step 1 Action * IMPORTANT Log out of the BTS prior to perform this procedure. Using a text editor, verify the following fields are set correctly in the bts–#.cdf file: EXAMPLE: 3 Asu1Equip = 1 Asu2Equip = 0 (1 if system is non-duplexed) Mc1Equip = 0 Mc2Equip = 0 Mc3Equip = 0 Mc4Equip = 0 RfdsEquip = 2 TestOrigDN = ’123456789’ TsuEquip = 1 NOTE The above is an example of entries extracted from the bts–#.
RFDS Setup and Calibration – continued Table 3-46: RFDS Parameter Settings Step 3 Action 5g – Click OK to close the status report window. 5h – Click on the MGLI. 5i – Click on Device in the BTS menu bar, and select Enable from the pull–down menu. –– A status report window is displayed showing status of the operation. 5j – When the operation is complete, click OK to close the status report window. ! CAUTION When the MGLI changes to INS_ACT, data will automatically be downloaded to the RFDS.
RFDS Setup and Calibration – continued Table 3-46: RFDS Parameter Settings Step Action – Click OK to close the status report window. 8c * IMPORTANT If the LMF displays an error message, check the following: Ensure AMR cable is correctly connected from the BTS to the RFDS. 3 Verify RFDS has power. Verify RFDS status LED is green. Verify entries in RFDS fields of the bts–#.cdf file are correct (refer to step 1).
RFDS Setup and Calibration – continued Table 3-47: Definition of Parameters IMSI MCC IMSI 11 12 These fields are obtained at the OMC using the following command: OMC000>disp bts–# imsi If the fields are blank, replace the IMSI fields in the NAM file to 0, otherwise use the values displayed by the OMC. MIN Phone Number These fields are the phone number assigned to the mobile. The ESN and MIN must be entered into the switch as well. 3 NOTE This field is different from the TODN field in the bts–#.
RFDS Setup and Calibration – continued Set Antenna Map Data The antenna map data must be entered manually if an RFDS is installed. Antenna map data does not need to be entered if an RFDS is not installed. The antenna map data is only used for RFDS tests and is required if an RFDS is installed. Prerequisite LMF is logged into the BTS 3 Follow the procedure in Table 3-49 to set antenna map data for the RFDS.
RFDS Setup and Calibration – continued Set RFDS Configuration Data If an RFDS is installed, the RFDS configuration data must be manually entered. Prerequisite LMF is logged into the BTS IMPORTANT 3 * The entered antenna# index numbers must correspond to the antenna# index numbers used in the antenna maps. Follow the procedure in Table 3-50 to set RFDS configuration data. Table 3-50: Set RFDS Configuration Data Step Action 1 Click on Util in the BTS menu bar, and select Edit > RFDS Configuration..
RFDS Setup and Calibration – continued RFDS Calibration The RFDS Calibration option is used to calibrate the RFDS TX and RX paths. TX Path Calibration – For a TX antenna path calibration the BTS XCVR is keyed at a pre–determined power level and the BTS power output level is measured by the RFDS. The power level is then measured at the TX antenna directional coupler by the power measuring test equipment item being used (power meter or analyzer).
RFDS Setup and Calibration – continued Test equipment and test cables are connected for TX calibration. Antenna map data has been entered for the site. BBXs are INS_TEST. Follow the procedure in Table 3-52 to perform RFDS calibration. Table 3-52: RFDS Calibration Step 3 Action 1 In the LMF, select the FRAME tab. 2 If it is not selected (no black dot showing), click on the B button in the BTS menu bar to select it.
RFDS Setup and Calibration – continued Program TSU NAM The NAM must be programmed before it can receive and process test calls, or be used for any type of RFDS test. Prerequisites MGLI is INS_ACT (bright green). SUA is powered up and has a code load. 3 Follow the procedure in Table 3-53 to program the TSU NAM. Table 3-53: Program NAM Procedure Step Action 1 In the LMF, select the RFDS tab. 2 Select the SUA by clicking on it.
Alarms Testing Alarm Verification The alarms testing should be performed at a convenient point in the optimization/ATP process, since the LMF is necessary to ensure that the RF cabinet is receiving the appropriate alarms from the power cabinet. The SC 4812ET is capable of concurrently monitoring 10 customer defined input signals and four customer defined outputs, which interface to the 50–pin punchblock. All alarms are defaulted to “Not Equipped” during ATP testing.
Alarms Testing – continued Heat Exchanger Alarm Test Table 3-54 gives instructions on testing the Heat Exchanger alarm. Table 3-54: Heat Exchanger Alarm Step Action 1 Turn circuit breaker “B” of the Heat Exchanger circuit breakers OFF. This will generate a Heat Exchanger alarm, ensure that the LMF reports the correct alarm condition in the RF Cabinet. 2 Alarm condition will be reported as BTS Relay #25 – “Heat Exchanger Alarm” makes contact. 3 Turn the circuit breaker “B” ON.
Alarms Testing – continued Minor Alarm Table 3-57 gives instructions on testing minor alarm. Table 3-57: Minor Alarm Step Action 1 Turn the Temperature Compensation Panel (TCP) power switch OFF. This will generate a minor alarm. Verify that the minor alarm LED (amber) is illuminated on the Meter Alarm Panel and the LMF reports this minor alarm. 2 Alarm condition will be reported as BTS Relay #24 “Minor Alarm” makes contact. 3 Turn the TCP power switch ON. The alarm condition should clear.
Alarms Testing – continued Multiple Rectifier Failure Table 3-59 gives instructions on testing multiple rectifier failure or major alarm in a three rectifier system. Table 3-59: Multiple Rectifier Failure or Major Alarm Step 1 Action With the rectifier module still in the unused shelf position fromTable 3-58 test procedures, turn the AC breaker for the 1st shelf OFF. 3 2 Verify that a rectifier alarm is generated.
Alarms Testing – continued Multiple Rectifier Failure (Six Rectifier System) Table 3-61 gives instructions on testing multiple rectifier failure or major alarm in a six rectifier system. Table 3-61: Multiple Rectifier Failure or Major Alarm Step 3 Action 1 Replace one rectifier module previously removed and turn the AC breaker for this shelf, OFF. 2 Verify that a rectifier alarm is generated.
Alarms Testing – continued Figure 3-24: Battery Overtemperature Sensor 3 Buss Bar FW00408 6 AWG Cables Battery Overtemp Sensor Negative Temperature Compensation Sensor Jan 2002 SC 4812ET BTS Optimization/ATP — CDMA LMF 3-113
Alarms Testing – continued Rectifier Over Temperature Alarm NOTE This is connector J8 on the rear of the Meter Alarm Panel itself, this is not connector J8 on the connector bulkhead at the rear of the cabinet. 3 Table 3-63 gives instructions on testing the battery over temperature alarm system. Table 3-63: Rectifier Over Temperature Alarm Step 1 Action Remove the J8 link on the rear of the Meter Alarm Panel (see Figure 3-25 for J8 location).
Alarms Testing – continued Figure 3-25: Location of Connector J8 on the Meter Alarm Panel FRONT VIEW VOLT AMP AMPS VOLT + + – PWR – TEST POINTS TEST POINTS OFF ON 3 REAR VIEW J1 J2 YEL VIOLENT OR J3 J8 J9 J6 J4 J5 Terminal Block RED BLK OR BRWN Terminal Block Rear Connector Panel J4 J5 Not Used J6 J1 J2 J3 FW00245 Before Leaving the site Table 3-64 gives instructions on what to check before leaving the site.
Alarms Testing – continued Notes 3 3-116 SC 4812ET BTS Optimization/ATP — CDMA LMF Jan 2002
Chapter 4: Automated Acceptance Test Procedure (ATP) Table of Contents Jan 2002 Automated Acceptance Test Procedures – Overview . . . . . . . . . . . . . . . . . . . . . Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ATP Tests Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TX/RX OUT Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ATP Test Procedure . . . . . . . . . . . . .
Table of Contents – continued Notes 4 SC 4812ET BTS Optimization/ATP — CDMA LMF Jan 2002
Automated Acceptance Test Procedures – Overview Introduction The Automated Acceptance Test Procedure (ATP) allows Motorola Cellular Field Engineers (CFEs) to run automated acceptance tests on all equipped BTS subsystem devices using the Local Maintenance Facility (LMF) and supported test equipment per the current Cell Site Data File (CDF) assignment. The results of these tests (at the option of the operator) are written to a file that can be printed.
Automated Acceptance Test Procedure – Overview – continued Reduced ATP NOTE Equipment has been factory–tested for FCC compliance. If license–governing bodies require documentation supporting SITE compliance with regulations, a full ATP may be necessary. Perform the Reduced ATP only if reports for the specific BTS site are NOT required.
Automated Acceptance Test Procedure – Overview – continued Full Optimization: Executes the TX calibration, download BLO, and TX audit before running all of the TX and RX tests. ATP Test Prerequisites Before attempting to run any ATP tests, ensure the following: BTS has been optimized and calibrated (see Chapter 3). LMF is logged into the BTS.
Automated Acceptance Test Procedure – Overview – continued ATP Test Procedure There are three different ATP testing options that can be performed to completely test a BTS. Depending on your requirements, one of the following ATP testing options should be run. Table 4-1 provides the procedure to execute an ATP test. To completely test a BTS, run the ATP tests according to one of the following ATP testing options.
Automated Acceptance Test Procedure – Overview – continued Individual Acceptance Tests The following individual ATP tests can be used to verify the results of specific tests: Spectral Purity TX Mask This test verifies that the transmitted CDMA carrier waveform, generated on each sector, meets the transmit spectral mask specification with respect to the assigned CDF file values.
Automated Acceptance Test Procedure – Overview – continued ATP Test Procedure Table 4-1 describes the step–by–step procedures to run any APT Test. Table 4-1: ATP Test Procedure Step Action 1 Select the device(s) to be tested. 2 From the Tests menu, select the desired test. 3 Select the appropriate carrier(s) (carrier – bts# – sector# – carrier#) displayed in the Channels/Carrier pick list. To select multiple items, hold down the or key while making the selections.
TX Spectral Purity Transmit Mask Acceptance Test Background: Tx Mask Test This test verifies the spectral purity of each BBX2 carrier keyed up at a specific frequency, per the current CDF file assignment. All tests are performed using the external calibrated test set, controlled by the same command. All measurements are through the appropriate TX OUT (BTS/RFDS) connector. The Pilot Gain is set to 541 for each antenna and all channel elements from the MCCs are forward-link disabled.
TX Spectral Purity Transmit Mask Acceptance Test – continued Figure 4-1: TX Mask Verification Spectrum Analyzer Display Mean CDMA Bandwidth Power Reference .
TX Waveform Quality (rho) Acceptance Test Background: Rho Test This test verifies the transmitted Pilot channel element digital waveform quality of each BBX2 carrier keyed up at a specific frequency per the current CDF file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements are via the appropriate TX OUT (BTS/RFDS) connector.
TX Pilot Time Offset Acceptance Test Background: Pilot Offset Acceptance Test This test verifies the transmitted Pilot channel element Pilot Time Offset of each BBX2 carrier keyed up at a specific frequency per the current CDF file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements will be via the appropriate TX OUT (BTS/RFDS) connector.
TX Code Domain Power Acceptance Test Background: Code Domain Power Test This test verifies the Code Domain Power/Noise of each BBX2 carrier keyed up at a specific frequency per the current CDF file assignment. All tests are performed using the external calibrated test set controlled by the same command. All measurements are via the appropriate TX OUT (BTS/RFDS) connector.
TX Code Domain Power Acceptance Test – continued Figure 4-2: Code Domain Power and Noise Floor Levels Pilot Channel PILOT LEVEL MAX OCNS CHANNEL 8.2 dB 12.2 dB MAX OCNS SPEC. Active channels MIN OCNS SPEC. MIN OCNS CHANNEL MAX NOISE FLOOR MAXIMUM NOISE FLOOR: < –27 dB SPEC. 4 Inactive channels Walsh 0 1 2 3 4 5 6 7 ... 64 Showing all OCNS Passing Pilot Channel PILOT LEVEL FAILURE – EXCEEDS MAX OCNS SPEC. 8.2 dB 12.2 dB MAX OCNS SPEC. Active channels MIN OCNS SPEC.
RX Frame Error Rate (FER) Acceptance Test Background: FER Test This test verifies the BTS FER on all traffic channel elements currently configured on all equipped MCCs (full rate at 1% FER) at an RF input level of –119 dBm [or –116 dBm if using Tower Top Amplifier (TMPC)]. All tests are performed using the external calibrated test set as the signal source controlled by the same command. All measurements will be via the LMF.
Generate an ATP Report Background Each time an ATP test is run, an ATP report is updated to include the results of the most recent ATP tests if the Save Results button is used to close the status report window. The ATP report will not be updated if the status reports window is closed with use of the Dismiss button.
Chapter 5: Leaving the Site Table of Contents Jan 2002 External Test Equipment Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 Reset All Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 Updating BTS CAL LMF Files in the CBSC . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1 BTS Site Span Configuration Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents – continued Notes 5 SC 4812ET BTS Optimization/ATP — CDMA LMF Jan 2002