User's Manual

(ESD)IEC 61000
-
4-2
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
input/output
lines
N/A
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000
-
4-5
1
kV differential
mode
2 kV common
mode
N/A
A Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
-
5 % UT(95% dip
in UT)
for 0.5 cycle, -40
% UT
(60 % dip in UT)
for 5 cycles, -70 %
UT(30 % dip in
UT) for 25 cycles,
-5 % UT
(95 % dip in UT)
for 5 sec
N/A
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the EQUIPMENT or SYSTEM requires
continued operation during power mains
interruptions, it is recommended that the
EQUIPMENT or SYSTEM be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
3 A/m
N/A
Power freq
uency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or of hospital environment.
DISPOSAL
Actuation of European directives 2002/95/EC, 2002/96/EC and 2003/108/EC, for reduction in use of
dangerous substances in the electric and electronic device and for garbage disposal. The symbol
applied on the device or its packaging means that at the end of its useful life the product must not be
disposed of with domestic waste. At the end of devices useful life, the user must deliver it to the able
collecting centers for electric and electronic garbage, or give back to the retailer when purchasing a
new device. Disposing of the product separately prevents possible negative consequences for the
environment and for health, deriving from inadequate disposal. It also allows the recovery of materials
of which it’s made up in order to obtain an important saving of energy and resources and to avoid
negative effects to the environment and health. In case of abusive disposal of device by the user, will
be applied administrative endorsements in compliance with current standard. The device and its parts
is marred with regard to disposal, as appropriate, in accordance with national or regional regulations.