Default Settings
The XXX System features Factory Default and User-Defined Default settings.
Factory Default Setting
In Factory Default setting, all adjustable alarm and volume parameters are set at
their default values. Factory Default setting is the system’s default operating setting.
It is indicated by DIP switch 8 in the DOWN position. For more information about
default alarm values, refer to this manual’s “Alarms and Limits” section.
User-Defined Default Setting
In User-Defined Default Setting (DIP switch 8 in the UP position), alarm limit
settings must be adjusted. When this setting is first activated, valid limit settings must
be entered for SpO
2
and pulse rate alarm limits; the system will not return to normal
operating mode until all limits have been set. Once set, the adjusted values are used
as defaults until the system is turned on with DIP switch 8 in the DOWN position,
at which time the device restarts in Factory Default setting.
IMPORTANT! In order to initiate User-Defined Defaults, the unit must
be turned on with DIP switch 8 in the UP position.