User's Manual
Table Of Contents
- Guide to Symbols
- Precautions for Use
- Using the XXX Pulse Oximetry System
- Indications for Use
- General Information about Bluetooth Technology
- Installing the Battery Pack in the Display Unit
- Installing Batteries in the Pulse Oximeter Device (POD)
- Displays, Indicators, and Controls
- Setting Up the XXX System
- Verifying System Operation
- Device Pairing
- Default Settings
- Accessing User Functions
- Dislay Unit DIP switches
- Care and Maintenance
- Alarms and Limits
- Communication
- Specifications
- Parts and Accessories
- Service, Support, and Warranty
- Troubleshooting
38
Troubleshooting
PATIENT OUT OF RANGE?
INTERFERENCE?
THE UNIT IS NOT PAIRED TO THE CORRECT POD
DETACH/REATTACH SENSOR
If these solutions do not correct the problem, please contact your distributor, or
contact NONIN Customer Support at (800) 356-8874 (USA and Canada) or
+1 (763) 553-9968.
The unit is in Alarm
mode, but no audible
alarms can be heard.
The 2-minute Alarm
Silence button is
activated.
Press the Alarm Silence button
to re-engage alarm volume, or
wait for two minutes—and
alarm tones will automatically
re-engage.
DIP switch 2 is in the UP
position, and the unit’s
volume is set to zero.
Adjust the alarm volume, or
return DIP switch 2 to the
DOWN position if you desire
audible alarms.
The system is not
functioning correctly.
Contact your distributor or
NONIN Customer Support.
The patient is out of
range.
Ensure that the patient stays
within the range of the XXX
System.
The devices will not
pair.
The sensor has not been
disconnected and
reconnected.
Disconnnect and reconnect
the sensor to the POD.
The patient is out of
range.
Ensure that the patient stays
within the range of the XXX
System.
Problem Possible Cause Possible Solution