Displays, Indicators, and Controls
11 Sensor Fault
This yellow indicator flashes when a sensor is disconnected, has failed,
or is not compatible with the monitor.
12 Poor Signal
This yellow indicator flashes when there has been a sustained period of
poor patient signals from the sensor.
Check the sensor site and reposition or replace the sensor if necessary.
13 Signal Processor Communication Error
This yellow indicator flashes and the message X-100SP not connected
displays when the respective signal processor has stopped
communicating with the display.
Check the signal processor connections or replace the signal processor
to correct the issue.
If the message appears in each channel, check the hub’s connection to
the monitor.
14 Alarm Silence
This yellow indicator flashes once every 2 seconds when the audible
alarm is silenced for 2 minutes.
If the alarm volume is at step 4 or lower (less than 45 decibels), the
Alarm Silence indicator is solidly lit.
15 Bluetooth
The Bluetooth indicator is green when Bluetooth is connected to a host,
white when it is enabled but not connected, and gray when it is disabled.
See "Connect/Disconnect Bluetooth" on page 51 for more information.
16
Full
Empty
l
Battery
The battery indicator shows the approximate percentage of battery life
remaining. When AC power is connected, the battery indicator fills up
repeatedly to indicate the battery is charging. The indicator stops filling
when the battery is fully charged.
Low – battery indicator flashes yellow
Critical – battery indicator flashes red
NOTE: When the X-100M reaches a low or critical battery condition, a
medium priority alarm sounds. To clear the audible alarm, plug in the
device. To clear the visible alarm, cycle power.
Table 3. X-100M Monitoring Screen Symbols and Indicators (Continued)
No. Symbol Description