Product Manual

18
Table 1: Limits Display Sequence
Parameter
Parameter
(SpO
2
)
Display
Initial
Setting
(Pulse Rate
Display)
Adjustment Range
Recall Alarm Settings rCL”“no”“yES” or “no
Low %SpO
2
Alarm Limit
02L
2,3
85” “Off”, 50 to 95 by 1
Pulse High Alarm Limit
HH
2
200” “Off”, 75 to 275 by 5
Pulse Low Alarm Limit
HL
2
50” “Off”, 30 to 110 by 5
High %SpO
2
Alarm Limit
02H
2
Off” “Off”, 80 to 100 by 1
Alarm Volume
adb
2
Hi” “Off” or “Lo” or “Hi
Clear Memory
Clr
1
no” “yES” or “no
Confirm Memory Clear
del
1
no” “yES” or “no
Year y” “00 0 to 99 by 1
Month nn” “00 0 to 12 by 1
Day d” “00 1 to 31 by 1
Hour h” “00 0 to 23 by 1
Minutes nn” “00 0 to 59 by 1
Notes:
1) Both of these menu options are part of the memory clear command; “del” will be
displayed only if yES” was selected as the setting for the “Clr” parameter.
2) These parameters are restored when Recall Alarm Settings is set to “yES.” These are also
the settings displayed by Review Alarm Settings.
3) The low SpO
2
Alarm limit saved for recall cannot be lower than the current default for that
alarm limit. If it is, the default value will be used when alarm limits are restored.