Product Manual

7
Displays, Indicators, and Controls
Front Panel Buttons
AC Power Supply LED
This green LED is displayed when an external power supply is providing
power to the Avant 9600. Any time this LED is displayed, the battery pack
is charging.
Battery LED
This amber LED indicates marginal battery capacity when blinking. When
solidly lit, the LED indicates that the battery pack is not installed. This
LED does not indicate that the device is running on battery power. New
battery packs must complete one charge/discharge/recharge cycle
before the Battery LED displays the actual capacity.
WARNING: The battery pack must be installed at all times while the device is
operating—even when operating on AC power. The audible alarms and memory will
not function if batteries are removed from the device. DO NOT use this device without
batteries.
ON/STANDBY
Press Once: Turns the device on
Hold 1 second: Turns the device off
Hold 5 seconds: Turns the device off when in Patient Security mode
Press momentarily while the unit is on:
Indicates the battery capacity in green, in 12% increments for
3 seconds
Initiates Print-on-Demand feature
Records event marker
Time/Date
This button displays the time and date. Year, month, day, hour, and minute
can be set using the Plus (+) and Minus (-) buttons.
Volume
This button allows users to set and review the pulse or alarm volume,
indicated by the designated LED.
Alarm Silence
Engages the 2-minute alarm silence automatically at start-up
Changes the alarm from silenced to audible
Silences the alarm for 2 minutes
Cancels audible and visual alarms, when in Locked Alarms mode, if
the alarm condition is no longer present.