Product Manual

24
Memory and Data Output Features
Patient Alarms
Only one alarm condition can be reported to the VueLink at a time. Because of this, patient
alarm conditions are assigned priorities—with only the highest priority alarm displayed at any
one time. Alarms are sent to the VueLink even when the oximeter’s alarms are silenced. Alarm
priorities are assigned as follows:
a. SpO
2
Low Limit
b. Pulse Rate High Limit
c. Pulse Rate Low Limit
d. SpO
2
High Limit
e. Low Perfusion
Equipment Alarms
VueLink displays equipment alarms as “inoperatives.” Only one inoperative condition can be
reported to the VueLink at a time. Because of this, inoperatives are assigned priorities—with
only the highest priority displayed at any one time. Inoperatives are sent to the VueLink even
when the oximeter’s alarms are silenced. Inoperative priorities are assigned as follows:
a. No SpO
2
Sensor
b. SpO
2
Sensor Fault
c. No SpO
2
Data
d. No Pulse Data
e. Low Battery
The following text strings, associated with the device, are sent through the VueLink module
for display on the multiparameter monitor:
Parameter Displayed Text
Displayed High SpO
2
Limit HiSpO
2
Displayed Low SpO
2
Limit LoSpO
2
Displayed High Pulse Rate Limit Hi PR
Displayed Low Pulse Rate Limit Lo PR
SpO
2
High Limit Alarm SpO
2
: HIGH
SpO
2
Low Limit Alarm SpO
2
: LOW
Pulse Rate High Limit Alarm PULSE: HIGH
Pulse Rate Low Limit Alarm PULSE: LOW
Low Perfusion Alarm LOW PERFUSION
Sensor Disconnect NO SpO
2
SENSOR