Product Manual

33
Technical Information
Table 11: Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
This device is intended for use in the electromagnetic environment specified below. The customer and/
or user of this device should ensure that it is used in such an environment.
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, relative
humidity should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV
±1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV
±2 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
±5% U
T
(>95% dip
in U
T
) for 0.5 cycle
±40% U
T
(60% dip
in U
T
) for 5 cycles
±70% U
T
(30% dip
in U
T
) for 25 cycles
<5% U
T
(>95% dip
in U
T
) for 5 sec.
±5% U
T
(>95% dip
in U
T
) for 0.5 cycle
±40% U
T
(60% dip in
U
T
) for 5 cycles
±70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 sec.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: U
T
is the AC mains voltage before application of the test level.