10-10 Height Sensor
To make large-scale adjustments to the probe tip height:
Only a trained service technician should perform this procedure.
1.
Raise the dispensing head so that neither the needle tip nor the probe tip are
touching the work surface.
2.
Using the 1/16 inch hex key, loosen the probe set screw accessed through the
hole on the right side of the Height Sensor (see Figure 10-1).
3.
Adjust the probe height by sliding the probe up or down (see Figure 10-3).
The recommended setting for the probe tip height is between 1.78 – 2.54 mm
(0.07 – 0.10 in) below the needle tip. However, visually aligning the probe
tip so that the bend in the probe is flush with the needle tip may be adequate
for your dispensing application. Consult Asymtek Technical Support for
further assistance.
4.
Tighten the probe set screw and adjust the probe position.
5.
Run a Height Sensor Setup procedure (see “Setup” on page 10-4).
Figure 10-3 Probe-to-Needle Alignment
Needle
1.78 - 2.54 mm (0.07 - 0.10 in)
1 mm (0.04 in)
H
eight Sensor
P
robe