System Configuration and Setup 4-37
DI-104 Scale
If your dispensing system is equipped with a DI 104 model scale, this dialog box will have a DI-104 tab
shown in Figure 4-20.
Figure 4-20 Setup Scale - DI-104
Calibration
Enter a target weight in the Calibration Weight text box. Place a calibrated weight of equal value in the
scale cup, then click on
Calibrate to calibrate the scale.
Environmental Conditions
The DI-104 scale includes several features designed to compensate for environmental variations, such
as vibration, drafts, etc. These environmental settings, which are selected during scale calibration
process, include:
 Filter - Very Low, Low, Normal, High, Very High
 Stability Speed - Very Slow (4 sec), Slow (2 sec), Normal (1 sec), Fast (0.5 sec), Very Fast
(0.25 sec)
 Stability Sensitivity - Very Fine (0.25 counts), Fine (0.5 counts), Normal (1 count), Coarse (2
counts), Very Coarse (4 counts)