Owner's Manual

36
Count
Ht:
Exit
Oset
Rel.
Hm:
HR:
120°2000
120°2000
[1]
120°2000
Aim at point A again by horizontal motion and tangent
screw and press [F3](Rel.);
Count
Ht:
Exit
Oset
Hold
Hm:
HR:
120°2000
120°0930
[ 1]
120°2000
Aim at point B again by horizontal motion and tangent
screw and press [F3](Hold);
Count
Ht:
Exit
Oset
Rel.
Hm:
HR:
240°4000
120°1800
[ 2]
120°2000
Repeat the steps above until the measurement times wanted;
If want to exit this function ,just press [F2](Exit),then press
[F4](Yes) to return the basic measurement;
Notice: if the angle value compared with the fist
measured is over ±30″, the error massage will appear.
5.6 Slope (V%)
Function: vertical angle switched with slope (V%).
Every time you press [F3] (V%),the display mode of vertical
angle will be switched each other. When the slope is over 45°it will
tip ―Over! ‖(Over measurement range).
5.7 H-Bz
When horizontal angle is in the range of (90°,180° or 270°)
± 4′30″, the buzzer will beep.