d. Saveairowtomemory
There is an ability to save a selected airow to memory so the device will revert
to the last known speed on power-on. The last known speed can be saved to
memory by holding the ON/OFF (1) and ‘-‘ (3) buttons for three seconds. To
revert to factory settings and to clear the air speed memory the ON/OFF (1) and
+’ (5) buttons are held for three seconds.
e. HEPAFilterreplacementtest
The HEPA lter replacement test indicator is for information purposes only.
The lter replace button, for testing the lter back pressure, is on the furthest
right-hand side of the panel. Upon pressing the lter back pressure test, the
device shall override the air ow setting to speed 5 air ow for testing the
lter for a period of approx. 30 seconds. The test button shall remain backlit
illuminated blue LED while the device is testing the air pressure.
The device will indicate that the HEPA lter needs to be inspected for
replacement by illuminating the test button in orange or a continuous LED light in
orange.
The device will indicate that the HEPA lter does not need to be inspected by
illuminating the test button in green or a continuous light in green.
Novaerus recommends replacing the HEPA lter every twelve-month (12) period
to ensure optimal performance, or when the lter test check output indicates to
do so; whichever comes rst. Discoloration on the HEPA lter may indicate that
the lter has exceeded its life expectancy. The expected lifetime of these lters is
approximately twelve months (12) under normal clean air conditions (such as that
of a hospital ward).