Manual
Controller N1100 
NOVUS AUTOMATION  7/8 
EVENT ALARM 
This function makes possible to program the activation of alarms in 
specific segments of a program. 
For such,  alarms  must have  their  function set  as  “rS
rSrS
rS  “ and be 
programmed in “ PE1
PE1PE1
PE1“ to “ PE7
PE7PE7
PE7“ according to Table 6. The number 
programmed in the event prompt defines the alarms to be activated. 
To configure and execute a ramp and soak program: 
•  Program the tolerance value, SV, time and event. 
•  If any event alarm is required program the ramp and soak event 
function. 
•  Set the control mode to automatic. 
•  Select  ramp and soak program to be executed at prompt  Prn
PrnPrn
Prn 
(0 to 7) 
•  Start control at the rvn
rvnrvn
rvn prompt by selecting YES. 
Before executing the program the controller waits for PV to reach the 
first set point SP0 if PtoL
PtoLPtoL
PtoL is different than zero. 
Should  any  power  failure  occur  the  controller  resumes  at  the 
beginning of the segment it currently is. 
CODE  ALARM 1  ALARM 2  ALARM 3  ALARM 4 
0
00
0 
1
11
1 
X       
2
22
2 
  X     
3
33
3 
X  X     
4
44
4 
    X   
5
55
5 
X    X   
6
66
6 
  X  X   
7
77
7 
X  X  X   
8
88
8 
      X 
9
99
9 
X      X 
10
1010
10 
  X    X 
11
1111
11 
X  X    X 
12
1212
12 
    X  X 
13
1313
13 
X    X  X 
14
1414
14 
  X  X  X 
15
1515
15 
X  X  X  X 
Table 5 - Event codes for ramp and soak 
AUTO TUNE 
During auto tune the process is controlled in ON / OFF mode at the 
programmed  SetPoint  (SV).  Depending  on  the  process 
characteristics large oscillations above and below SV may occur and 
auto tuning may take several minutes to be concluded. 
The recommended procedure is as follows: 
•  Disable the control output at the rvn
rvnrvn
rvn prompt by selecting NO. 
•  Select auto mode operation at the Avto
AvtoAvto
Avto prompt by selecting YES. 
•  Disable the ramp and soak function (select NO) and program a new 
SV value other than the present PV (close to the desired set point). 
•  Enable auto tuning at the Atvn
AtvnAtvn
Atvn prompt by selecting YES. 
•  Enable the control output at the rvn
rvnrvn
rvn prompt by selecting YES. 
During  the  auto tune  procedure the soft-start  function will not operate 
and large oscillations will be induced around the setpoint. Make sure the 
process can accept these oscillations and fast control output changes. 
If auto tuning results are not satisfactory refer to Table 6 for manual 
fine tuning procedure. 
PARAMETER  RESPONSE  SOLUTION 
Slow Response  Decrease 
Proportional Band 
Large Oscillation  Increase 
Slow Response  Increase 
Integral Rate 
Large Oscillation  Decrease 
Slow Response or Instability  Decrease 
Derivative Time 
Large Oscillation  Increase 
Table 6 - Suggestions for manual tuning of PID parameters 
PROBLEMS WITH THE CONTROLLER 
Connection  errors  and  inadequate  programming  are  the  most 
common errors found  during  the  controller  operation.  A  final review 
may avoid loss of time and damages. 
The controller displays  some messages to help  the user identify 
problems. 
MESSAGE  PROBLEM 
----
--------
---- 
Open input. Without sensor or signal. 
Err1
Err1Err1
Err1 
Err6
Err6Err6
Err6 
Configuration or connection problem in the Pt100 
cable 
Other error messages displayed by the controller can account for errors 
in the input connections or type of selected input non compliant with the 
sensor  or  signal  applied to  the  input.  If  errors  persist, even after a 
review, contact the manufacturer. Inform also the device serial number. 
To find out the serial number, press 
 for more than 3 seconds. 
The controller also has a visual alarm (the display flashes) when the 
PV value is out of the range set by spxl
spxlspxl
spxl and spll
spllspll
spll. 
CALIBRATION 
INPUT CALIBRATION 
All inputs are factory calibrated and recalibration should only be done 
by qualified personnel. If you are not familiar with these procedures 
do not attempt to calibrate this instrument. The calibration steps are: 
a)  Select the input type to be calibrated. 
b)  Set the desired upper and lower display limits. 
c)  At the input terminals inject an electrical signal corresponding to a 
known indication value a little higher than the lower display limit. 
d)  Select theinL(
inL(inL(
inL( prompt. Through the 
 and 
 keys adjust PV 
so that it matches the injected signal. 
e)  Inject a signal that corresponds to a value a little lower than the 
upper limit of the display. 
f)  Select theink(
ink(ink(
ink( prompt. Through the 
 and 
 keys adjust PV 
so that it matches the injected signal. 
g)  Repeat steps c) to f) to improve calibration. 
ANALOG OUTPUT CALIBRATION 
1)  Select type 11 or 12 at the I/O5 prompt. 
2)  Connect a current meter at the analog output. 
3)  Disable the auto-tune and soft-start functions. 
4)  Set the output low limit ovLL
ovLLovLL
ovLL to 0.0 % and the output high limit 
ovkL
ovkLovkL
ovkL to 100.0 %. 
5)  Select the manual mode at the avto
avtoavto
avto prompt. 
6)  Enable the output at the rvn
rvnrvn
rvn prompt. 
7)  At the operation cycle, set the MV to 0.0 %. 
8)  At the output low calibration ovLC
ovLCovLC
ovLC prompt, press the 
 and 
key until the mA meter reads zero mA. Approach this value from 
above. 
9)  Set 100.0 % for the manipulated variable (MV). 
10)  At  the  output  high  calibration  ovkC
ovkCovkC
ovkC  prompt,  press  the 
  and 
  key  until  the  mA  meter  reads  20  mA.  Approach  this  value 
from below. 
11)  Repeat steps 7) to 10) as necessary. 








