Manual
myPCLab 
NOVUS AUTOMATION  2/9 
  Timing: It times up to 4294967295 miliseconds (more than 49 days). It is selectable to run timing when 
input in logic level "0" or when in "1". Accuracy: 0.5 % of the indicated time. 
  Frequency: It  counts  the  number  of pulses  (rising edge, falling  edge  or  both)  in each base  time, 
configurable from 1  to 65535 seconds  (18 hours). Maximum  input frequency: 500  Hz (square wave, no 
debounce). Maximum input frequency: 500 Hz (square wave, no debounce). Accuracy: 0.5 %. 
Ambient Temperature Sensor: Internal thermistor. 
Total Accuracy: Thermocouples R, S and B: 0.25% of the maximum range ± 3 °C (with maximum A/D resolution); 
          Other thermocouples: 0.25% of the maximum range ± 1 °C (with maximum A/D resolution); 
          Pt100, voltage and current: 0.20% of the maximum range (with maximum A/D resolution); 
          Ambient temperature channel: ± 1.5 °C (after 20  minutes connected to the USB port). 
A/D Resolution: Configurable from 15 to 11 bits. 
Input Sampling Rate:   From  8  to  128  samples  per  second,  depending on  configured  A/D  resolution  and  number  of 
enabled channels. 
Computer Interface: USB V1. 1 Plug and Play, virtual serial port interface. 
USB connection: Mini-B receptacle. 
Power supply: From the USB bus. Typical current 30 mA. 
Operating environment: 0 to 50°C, 10 to 90% relative humidity, non-conden sing. 
Electromagnetic Compatibility: EN 50081-2, EN 50082-2. 
Built-in cold junction compensation for thermocouples and wire resistance for Pt100. 
ABS case, dimensions: 70 x 60 x 18 mm. See figure. 