Manual

Never stack adapters to measure levels
higher than +40dBu!
ML1
Adapter -20dB
Thank you for buying the ML1 -20dB adapter! This
device extends the XLR input range of your Minilyzer
up to +40dBu.
Use the adapter only in conjunction with the Minilyzer.
The impedances of the adapter are machted for this
use only!
!
Never connect the adapter to mains
voltage!
!
Usage:
Connect the adapter directly to the XLR input of the
Minilyzer, as shown on the picture.
Technical Specications:
Attenuation: 20 dB
Initial accuracy: +/- 0.05dB (aging < 0.02dB @1000h full load)
Input impedance: Bal.: 40kOhm Unbal.: 20kOhm
Balance missmatch: < 0.2%
Max. input voltage: Bal.: +40dBu 77.5VRMS 218Vpp
Unbal.: +34dBu 38.9VRMS 108Vpp
www.minstruments.com