(5) When the measurement is completed, the
acceleration and its frequency of the top 10 of
the FTT analysis result, as well as judgment
result based on the set judgment criterion
value, assumed damaged position, OA
measurement result value and the judgment
result based on the set judgment criterion
value, ISO-10816 judgment result are
displayed.
Figure 4.32 Result screen of Measure with
Meas. Point
Table 4.2 FFT judgment
Condition Judgment
80% or less of
the
judgment
criterion
value
Good
E
xceeding 80%
and
100% or less
of the judgment
criterion
value
Caution
Exceeding th
e judgment
criterion
v
alue
Warning
Table 4.3 OA judgment
Condition Judgment
Less than
the
judgment
criterion
value
Measurement
value
is
displayed
in blue
.
E
xceed
ing t
he judgment
criterion
value
Measurement
value
is
displayed
in red
.
Table 4.4 ISO-10816 judgment
Judgment
Good
Acceptable
Caution
Warning
(6) When the "FFT" button on the bottom left of the
measurement result screen is tapped, FFT
spectrum is displayed. When the frequency
button at the bottom of the graph is tapped, the
frequency range of the graph can be specified.
When "Time Waveform" on the upper right of
the screen is tapped, the time waveform is
displayed.
Figure 4.33 FFT spectrum screen of Measure
with Meas. Point
(7) By using the bottom left button on the graph,
raw waveform or the waveform after envelope
processing can be selected.
Figure 4.34 Time waveform screen
(raw waveform)