Guidance and manufacturer’s declaration electromagnetic immunity
The THixyz series is intended for use in the electromagnetic environment specified below.
The customer or the user of the THixyz series should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of the THixyz series, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz
to
80 MHz
Not
applicable
Recommended separation distance
d = 1,2
d = 1,2 80 MHz to 800 MHz
d = 2,3 800 MHz to 2,5 GHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to
2,5 GHz
3 V/m
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a
should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity
of equipment marked with the
following symbol: