Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual
29
Indicates an electrode. Gray is inactive, yellow is a cathode, blue is an anode.
Percentage is the percentage of the subprogram amplitude allocated to the electrode.
e electrode number is in the lower right.
Electrode polarity is in the lower le.
e can displays at the bottom right of the right-most lead. You can program the can like an
electrode.
Port Indicator. e port (lead) on which the program stimulates. For a three bore stimulator, red
is port 1, yellow is port 2, and green is port 3. For a two bore stimulator, red is port 1 and green is
port 2.
Note: Colors indicate port location and may not match ags used during the implant procedure.
Undo any unsaved changes.
e name of the program. Tap to change the program name. e program name is visible to the
patient on the Patient Programmer Charger.
e name of the subprogram. Tap to change the subprogram name.
e number of the subprogram. Tap to congure a subprogram.
Locks a selected electrode from automatic amplitude adjustments.
Tap the electrode then tap the lock to lock the selected electrode. A small lock icon displays next to
the locked electrode. To unlock the electrode, tap the locked electrode and tap the lock again.
e currently selected electrode.
e subprogram amplitude allocation applied to the selected electrode or can. Tap arrows to adjust
the percentage of amplitude allocated to the electrode.
Note: Adjusting the percentage allocated to the electrode also adjusts the amplitude allocated to
other assigned electrodes that are unlocked. A subprogram must have at least two electrodes to
adjust the amplitude allocation.
Removes the selected electrode or can from the subprogram.
Sets the polarity of the electrode or can.
Controls stimulation for the current subprogram or entire program. e blinking green light
indicates stimulation is on. e blinking green light turns o to indicate stimulation is o.
Tap ON or OFF to turn stimulation on or o. Select Subprogram or Program. Upon starting
stimulation, stimulation ramps to the amplitude value for the subprogram. When stimulation is on
and you are switching between subprograms, the stimulation for the currently selected subprogram
ramps down before switching and ramping the stimulation for the newly selected subprogram.
0300-000176-001 Rev.2 SCS CP_Mobile.indb 29 7/12/19 9:20 AM