Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual
31
5. Tap the Subprogram tab.
6. Set the stimulation parameters
(amplitude, pulse width, and frequency) for the subprogram by tapping on the le and
right arrows for each stimulation parameter
.
Note: Adjusting the frequency for a subprogram adjusts the frequency for all subprograms within the program.
e amplitude of the pulse in milliamps (mA).
e amplitude for an electrode cannot exceed 15 mA.
e total amplitude for a subprogram cannot exceed 30 mA.
e amplitude for a subprogram can be set from 0–15 mA for one electrode per polarity,
or 15–30 mA for two or more electrodes per polarity.
e pulse width in microseconds (μs).
e pulse width range is 20–1500
μs.
e frequency of the pulse in Hertz (Hz).
e frequency range is 2–2000 Hz.
7. Adjust the stimulation parameter ranges. e range also sets the ranges that the patient can adjust.
a. Tap the displayed amplitude (mA), pulse width (μs), or frequency (Hz) value
on the Subprogram
tab.
b. e range slider popup displays.
c. Tap and drag the slider dots to set the minimum and maximum values.
Note: For amplitude, you can adjust the step. e step is the rate at which the amplitude value increases or decreases
when tapping the arrows.
d. Close the popup.
8. Tap Summary on the Navigation menu to advance to the Summary screen and save program settings.
Close
Range slider dots
Step size
0300-000176-001 Rev.2 SCS CP_Mobile.indb 31 7/12/19 9:20 AM