32 Clinician Programming Manual
Shifting the Amplitude
Shifting the Amplitude
Shiing moves the amplitude allocation between and across leads.
To shi the electrode conguration:
1. Tap the Shiing tab on the Programming screen.
2. Tap the desired rate for the Shiing Increment.
3. Use the up, down, le, or right arrows to shi the amplitude allocation.
Note: Shiing arrow availability is determined by lead placement.
Optionally, lock an electrode from shiing by tapping the electrode on the Programming screen. A small blue lock icon
displays on the locked electrode.
Note: e shiing lock is dierent from the subprogram automatic amplitude adjustment lock.