Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual
43
10. Stimulator Settings
Use the Stimulator Settings to view information about the stimulator, congure background impedance checks, and turn
on advanced patient adjustments.
Accessing Stimulator Settings
To access stimulator settings:
1. Select a patient from the Patient Directory.
2. Select Stimulator Settings from the Navigation menu.
View battery information. Refer to Stimulator Battery on page 44.
View a diagnostic message generated by the stimulator.
Congure background impedance checks. Refer to Impedance Check on page 46.
Optimize communication between the Patient Programmer Charger and the stimulator. Refer to Auto
Tune on page 45.
Restart the stimulator, which stops stimulation and reinitializes the internal settings.
Note: is does not aect the patient information or programs on the stimulator.
Turn magnet mode on or o. When on, the patient can use a magnet to stop and start stimulation. Refer
to the Patient Magnet Manual for more information.
Note: Regardless of this setting, you can always use a magnet to stop stimulation.
Place the stimulator in storage mode.
Use the Patient Programmer Charger to take the stimulator out of storage mode.