Programming Manual

56 Clinician Programming Manual
Charging a Fully Discharged Implantable Stimulator
Charging a Fully Discharged Implantable Stimulator
A patient may need assistance to charge a fully discharged implantable stimulator.
To charge a fully discharged implantable stimulator:
1. Move the Patient Programmer Charger within 1 meter (3 feet) of the implantable stimulator and turn on the Patient
Programmer Charger. e Not Able to Connect screen displays.
2. Tap Cancel to display the Home screen.
3. Connect the charging paddle to the Patient Programmer Charger.
4. Tap
(Charge Status). e Charge Status screen displays.
5. Align the charging paddle over the implantable stimulator. Use either the adjustable belt or an adhesive patch to hold
the charging paddle in place.
6. Tap OK on the Charge Status screen.
7. Wait until charging completes.
8. Tap OK on the Charging Complete screen.
9. Tap
(Communication) to connect to the implantable stimulator.
Setting the Implantable Stimulator Depth
Adjust the depth setting if an implantable stimulator is implanted deeper than 1.5 cm (0.59 in) or if the patient has trouble
charging the implantable stimulator.
To set the implant depth:
1. Tap
(Programmer Settings) on the Home screen of the Patient Programmer Charger.
2. Tap Advanced Settings on the Programmer Settings screen.
3. Enter the last two digits of the serial number located on the back of the Patient Programmer Charger, and tap OK.
4. Tap Implant Depth on the Advanced Settings screen.
5. Select Deep.
0300-000176-001 Rev.2 SCS CP_Mobile.indb 56 7/12/19 9:20 AM