58 Clinician Programming Manual
Bridge Communicator Specifications
Guidance and manufacturers declaration – electromagnetic emissions
e Bridge Communicator is intended for use in the electromagnetic environment specied below. e customer or the user of the Bridge
Communicator should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 e Bridge Communicator uses RF energy primarily for its internal function. erefore, its RF
emissions are very low and are not likely to cause interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B e Bridge Communicator is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/
icker emissions
IEC 61000-3-3
Complies
Guidance and manufacturers declaration – electromagnetic emissions
e Bridge Communicator meets basic safety requirements for use in the electromagnetic environment specied below. e customer or the user of
the Bridge Communicator should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete, or ceramic tile. If oors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV dierential
mode
± 2 kV common
mode
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 s
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% UT
(30% dip in U
T
)
for 25 cycles
<5% UT
(>95% dip in U
T
)
for 5 s
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Bridge Communicator requires continuous
operation during power mains interruptions, it is recommended that the Bridge
Communicator be powered from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.