Programming Manual

Algovita Spinal Cord Stimulation System
Clinician Programming Manual
7
1. Introduction
About this Manual
is Clinician Programming Manual provides instructions on using the Algovita® Clinician Programmer Application and
on using, charging, and caring for the Bridge Communicator. e Algovita® Spinal Cord Stimulation (SCS) System delivers
mild electrical pulses to a specic area in the spinal cord. Patients may feel stimulation as a tingling sensation (paresthesia)
that masks or covers their pain.
is manual applies to Model 4510 Algovita Clinician Programmer Application and Model 4110-xx Bridge Communicator
for use with Algovita Stimulator Models 2408 and 2412, and Algovita Trial Stimulator Model 4300.
Clinicians use the Clinician Programmer intraoperatively to check system integrity and determine lead placement. ey
use it postoperatively to evaluate and program stimulation therapy, and to associate the Bridge Communicator, Patient
Programmer Charger, and Pocket Programmer with a stimulator.
Spinal Cord Stimulation System
e Algovita SCS System is a rechargeable, 24-electrode, SCS system for the treatment of chronic pain. e implanted
components of the SCS system consist of an implantable stimulator and leads with optional extensions. During a
stimulation trial, trial leads are placed percutaneously, with the end of the trial lead externalized and connected to a trial
stimulator. e trial stimulator replaces the implantable stimulator during intraoperative test stimulation and during the
stimulation trial.
Note: During stimulation trials, you can use a commercially available ground pad with the Algovita SCS System.
1
2
1
2
3
Patient Programmer Charger
with Charging Paddle
Clinician
Programmer
Model 4500
Clinician Programmer
Model 4510 with Bridge
Communicator Model 4110-xx
Pocket
Programmer
Implantable
Stimulator with Leads
Trial
Stimulator
0300-000176-001 Rev.2 SCS CP_Mobile.indb 7 7/12/19 9:20 AM